© Edwards Limited 2015. All rights reserved.
Page 9
Edwards and the Edwards logo are trademarks of Edwards Limited.
Technical data
D396-20-880 Issue H
2
Technical data
2.1
Operating and storage conditions
2.2
Mechanical data
2.3
Electrical data
Table 1 - Operating and storage conditions
Ambient operating temperature range
0 to 40 °C
Ambient storage temperature range
-20 to 70 °C
Maximum ambient operating humidity
10 to 95% RH (non-condensing to DIN 40040)
Maximum operating altitude
3000 m
Cooling
Natural convection
Table 2 - Mechanical data
Dimensions
See
Mass
EXC100E
0.8 kg
EXC100L
1.0 kg
Enclosure protection
IP20
Pollution degree
EN61010, category 2
Table 3 - Electrical data
Electrical supply
Voltage
90 to 264 V a.c., single-phase
Frequency
47 to 63 Hz
Maximum input power
220 VA
Peak inrush current
11 A at 110 V a.c.
40 A at 240 V a.c.
Fuse rating
2 A, type T 20 mm
Over-voltage transients
EN61010, category 2