5. Technical data
Table 4 Technical data
EDS
200
300
Pumping speed without gas-ballast
> 210 m
3
/h
> 280 m
3
/h
Ultimate partial pressure
< 0.05 mbar
< 0.01 mbar
Maximum permissible inlet pressure
1013 mbar
Max. permanent discharge pressure
(absolute pressure)
1200 mbar
Swept volume maximum pressure rating
11 bar
Integral leak rate
< 10
-4
mbar l/s
Water vapour capacity without
condensation between 10 – 50 mbar
1.5 - 7 kg/h
2 - 10 kg/h
Permissible ambient temperature
5 to 40 °C
Storage temperature
-30 to +50°C
Typical continuous A-weighted sound
pressure level at ultimate pressure
(without silencer but with long exhaust
line)
≤ 72 dB(A)
Relative atmospheric humidity
max. 90%
Installation location
up to 2000 m (m ASL)
Cooling
Water and water/ glycol mixture
Mains voltage
200 - 230 V or 380 - 460 V AC ± 10% or 575 V
Frequency
50/60 Hz
Phases
3-ph
Motor rating
5.5 kW (7.5 HP)
7.5 kW (10 HP)
Current rating (200 – 230 V systems) @
full load
23 A
29 A
Current rating (380 – 460 V systems) @
full load
11.5 A
14.5 A
Recommended branch circuit fuse:
UL (200 – 230 V systems)
IEC (200 – 230 V systems)
UL (380 – 460 V systems)
IEC (380 – 460 V systems)
UL (575 V systems)
IEC (575 V systems)
32 A
32 A
16 A
16 A
16 A
16 A
Min cable size for:
200 – 230 V systems
380 – 460 V systems
575 V systems
6 mm
2
4 mm
2
4 mm
2
Power consumption at ultimate pressure
4.1 kW (5.5 HP)
4.5 kW (6 HP)
Installation category
II (IEC 60664-1)
Input supply voltage unbalance
Should not exceed 2% when assessed over any one
minute period
Page 17
A41802880_C - Technical data
Содержание EDS 200
Страница 48: ...Figure 21 EDS 200 300 with water cooling dimensions in mm Page 48 A41802880_C Dimension drawings...
Страница 49: ...13 Wearing parts Gasket for plug screw G 3 8 oil fill plug ES23955165 Page 49 A41802880_C Wearing parts...
Страница 53: ...This page has been intentionally left blank...
Страница 54: ...This page has been intentionally left blank...
Страница 55: ...This page has been intentionally left blank...
Страница 56: ...edwardsvacuum com...