Maintenance
D397-21-880 Issue G
5
Maintenance
5.1
Safety
There are no serviceable parts on the TIC. Do not open, return to your nearest Edwards service centre for any repairs
that are necessary.
The Edwards return of equipment forms can be found at the rear of this manual.
5.2
Fault finding
WARNING
Obey the safety instructions given below and take note of the appropriate precautions. If you do
not you cause injury to people and damage to equipment.
Table 22 - Fault finding
Diagnostic messages
Description
Ga
ug
e
Over Range
Refer to appropriate Section in manual.
Under Range
Refer to appropriate Section in manual.
New ID
Refer to
Uknw
Refer to
Not connected
Refer to
Not Struck
Refer to appropriate Section in manual.
Filament Fail
Refer to appropriate Section in manual.
Mag Fail
Refer to
Striker Fail
Refer to
Cal Error
Refer to
Initialising
Refer to
Emission Error
Refer to
Over Pressure
Refer to
ASG Cant Zero
Refer to
Gen
er
al
SYSI Inhibit
The system interlock has been disconnected. Please check that the logic
interface plug is connected correctly, or check the status of the system
interlocks.
Ext Inhibit
Enable lines have been disconnected, please check your external inhibit
lines.
No Reading
An object has not received a value update from its source within a given
time and is flagging that its value is now old. Check connections to
components of system.
No Message
Brownout/short
An object has not received a reply to a message it sent within a given time.
Check logic interface connections, are correctly attached to the TIC.
The loading on the TIC outputs exceeds a certain current level. Remove the
short or high load and then accept the error, the TIC will re-attempt to
drive the output.