B756-50-880 Issue C
Page 8
© Edwards Limited 2013. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
INSTALLATION
If the system is not to be used immediately, store it in suitable conditions as described in
. Do not discard
the packing materials; retain them to re-package the pump if it is returned for service.
3.2
Installation
CAUTION
Do not invert the system.
The EXT70/70/200H system is intended for installation on the Waters LCT Premier instrument only. Refer to the
customer system fitting instructions for further details.
Once the system has been removed from the packaging, the two eyebolts used to lift it can be unscrewed from their
positions. One of the supplied eyebolts should be temporarily fitted to the side position highlighted on
to
aid manoeuvring the system on to the LCT Premier flight tube.
Table 2 - List of items supplied
Qty
Description
Check (
)
1
System chamber
1
EXT70/70/200H Split Flow Turbomolecular pump (fitted to chamber)
1
Top Interface Plate (fitted to chamber)
1
Junction piece (fitted to chamber)
2
Side Inlet screen (fitted to pump)
2
133 x 3.0 section individually sealed O-ring seals
1
EXDC160 controller (fitted to pump)
*
*
Only supplied with whole systems. The cartridge pump-only option does not have EXDC160
controller fitted when supplied.
WARNING
Install the pump into the vacuum system before connecting the controller to the power supply.
This will ensure that the pump cannot operate and injure people during the installation.
WARNING
Ensure that connections to the pump do not generate slip or trip hazards.
WARNING
If the pump seizes, the stored energy can cause rapid movement of the pump, which may cause
further damage and injury to people. Ensure the EXT70/70/200H is securely fixed to the system
chamber, and that the system chamber is securely fitted to the LCT PREMIER application.