A313-01-880 Issue A
Page 22
© Edwards Limited 2008. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Installation
Figure 5 - Inlet/outlet flange loading limits (Nm)
A. Side view
B. Plan view
1. Inlet/outlet flange
2. Outlet/inlet flange
F Force
M Moment
R Resultant
X Load orientation
Y Load orientation
Z Load orientation
Force in plane
X
Y
Z
R
Inlet flange
1000 1000 2000 2400
Outlet flange
750
750
1500 1800
Moment in plane
X
Y
Z
R
Inlet flange
500
500
500
870
Outlet flange
300
300
300
250