Changing Tips
1.
Turn Tip counter-clockwise by using a
WT620
Socket Wrench.
2.
Remove and replace with desired Tip.
Hot Air Calibration
It is highly recommended to use new or very clean thermocouple
wires.
FOLLOW SET-UP AS ILLUSTRATED
1.
With the Hot Air Tip inside the
TPL09-1
Ceramic Temperature Sensor, place
the center of the thermocouple wire of the
SDS100
Temperature Sensor Dock
inside the slot of the
TPL09-1
.
2.
Push the power switch and set the Temperature Control Knob to 400°F.
3.
Turn Flow Regulator Knob to 4 - 5 SCFH.
4.
Adjust LO-Temp. Calibration Pot so the Meter will read 400°F.
5.
Set Temperature Control Knob to 800°F.
6.
Adjust Hi-Temp Calibration Pot so the Meter will read 800°F.
MS412
Digital Temperature
Measurement System
For a quicker calibration for more than one
station try using the
TP475
Temperature Probe
RCS75
WT620
Hot Air Tip
Use a 5/16 wrench
to hold the sleeve if
tip is too tight
TI680
SDS100
FX635
Hot Air Tip
TPL09-1