5.4
Field Exchangeable Frequency Operations
The Field Swappable functionality is only possible with 230 / 550 kHz Side Scan systems.
In order to switch between different bathymetry frequencies, the 6205s arrays must be exchanged with
the corresponding bathymetry array. For instance, if the Sonar Head is configured as a 230 kHz/550 kHz
Side Scan with 230 kHz bathymetry and the user wants to exchange the bathymetry frequency to the 550
kHz channel, then the following procedure should be followed.
Tools
•
1/16-inch Allen Wrench
•
3/16-inch Allen Wrench
•
Tube of Thread Lock
1.
Place the 6205s on a stable and level platform, such as a workbench or boat deck.
Figure 5-4: Unit on Stable Platform
2.
Removing the [8] cap screws using a
1
/
16
” Allen key on the right and left sides of the housing.