40 | www.eddyfi.com
Using Reddy with Sharck Probes
Figure 5–25
Sharck probe on carbon steel weld
5.
On Reddy’s keypad, press the Start/Stop acquisition button (see page 3).
Alternatively, in the front stage view, on the
Home
or
Calibration
ribbon, tap
Acquire
.
Note
To ensure that the entire weld cap is covered by the probe’s fingers, whenever possible, position the
center of the probe’s active area on the centerline of the weld. Depending on the width of the weld
crown and the welding process, the probe may also cover all or part of the heat-affected zone (HAZ).
6.
Placing appropriate pressure on the probe, move it longitudinally along the weld.
Note
The maximum speed is 200 mm/s (6 in/s) for the small (W028) and medium (W053) Sharck probes.
7.
When scanning is complete, on Reddy’s keypad, press the Start/Stop acquisition button (see page
Alternatively, in the front stage view, on the
Home
or
Calibration
ribbon, tap
Stop
.
Notes
•
To maintain data files at a reasonable size and facilitate analysis, when inspecting very long welds,
we recommend that you divide your inspection into several appropriately short acquisitions (e.g.
less than 3 m (10 ft)).
•
Balance the probe regularly between scans on the plastic portion of Eddyfi’s normalization plate
(REFPL-CAL-SHARCK-AAL-LG) to avoid any signal drift.
•
Data is positioned according to the origin (0,0) illustrated below. This is the top view of the
probe (transparency). Along the scanning axis, the origin is 38.75 mm (1.53 in) from the front of
the probe’s casing (the front being the edge of the casing closest to the scan direction).
Figure 5–26
Probe origin and defect positioning along the scanning axis (probe shown top view)
•
Data simultaneously appears on the Reddy display during acquisition and some of the functions
of the
Home
and
Calibration
ribbons are disabled.
Содержание Reddy
Страница 1: ...User s Guide...
Страница 9: ...ix...
Страница 10: ...x www eddyfi com General Precautions and Conventions General Precautions and Conventions...
Страница 15: ...xv General Precautions and Conventions...
Страница 16: ...2 www eddyfi com Reddy Overview Chapter 1 Reddy Overview...
Страница 27: ...13 Reddy Overview...
Страница 28: ...14 www eddyfi com MagnifiGO Overview Chapter 2 MagnifiGO Overview...
Страница 34: ...20 www eddyfi com MagnifiGO Overview Figure 2 6 Acquisition Summary Figure 2 7 Report Summary...
Страница 35: ...21 MagnifiGO Overview Figure 2 8 Save Figure 2 9 Material Database...
Страница 36: ...22 www eddyfi com Preferences Chapter 3 Preferences...
Страница 38: ...24 www eddyfi com Preferences Figure 3 3 Analysis preferences Figure 3 4 Help...
Страница 39: ...25 Preferences...
Страница 40: ...26 www eddyfi com Keypad and Keyboard Functions Chapter 4 Keypad and Keyboard Functions...
Страница 42: ...28 www eddyfi com Using Reddy with Sharck Probes Chapter 5 Using Reddy with Sharck Probes...
Страница 62: ...48 www eddyfi com Maintenance and Troubleshooting Chapter 6 Maintenance and Troubleshooting...
Страница 66: ...52 www eddyfi com Specifications Chapter 7 Specifications...
Страница 69: ...55 Specifications...
Страница 70: ...56 www eddyfi com Connector Reference Appendix A Connector Reference...
Страница 74: ...60 www eddyfi com Using the Optional Harness Appendix B Using the Optional Harness...
Страница 82: ...www eddyfi com...