| 39
Using Reddy with Sharck Probes
7.
In the
Sharck Assisted Calibration
dialog box, tap
Aluminum
.
When the necessary data is acquired, a green check mark appears next to
Aluminum
.
8.
Position the Sharck probe on the carbon steel component to be tested.
9.
Apply even pressure on the probe.
10.
In the
Sharck Assisted Calibration
dialog box, tap
Carbon steel
.
11.
Move the probe over a distance of 200 mm (7.9 in) on the component to be tested (over the weld).
You have approximately 4 seconds to carry out the scan from the moment you tap the
Carbon
steel
button. When the necessary data is acquired, a green check mark appears next to
Carbon
steel
.
Note
If non-magnetic paint or coating is present, there is no need to remove it (unless it is thicker than
3 mm (0.118 in). The Sharck calibration process takes this liftoff into account and manages accordingly.
12.
In the
Sharck Assisted Calibration
dialog box, tap
Calibrate
.
When the necessary data is acquired, a green check mark appears next to Calibrate.
Figure 5–24
Sharck Assisted Calibration: Calibration complete
Note
If, for any reason, you must normalize the probe again, tap
Reset
before you perform the above
procedure.
13.
To close the dialog box, tap the red X in its upper-right corner.
Collecting Data
1.
Position the Sharck probe on the plastic portion of the normalization plate (REFPL-CAL-SHARCK-
AAL-LG).
Note
The normalization plate should be on a steady, flat surface and shall be at least 10 cm (4 in) from any
ferritic surface.
2.
Apply even pressure on the probe.
3.
On Reddy’s keypad, press the Probe nulling button (see page 3).
Alternatively, in the front stage view, on the
Home
or
Calibration
ribbon, tap
Null
.
4.
Position the probe on the part to be inspected across the weld, as illustrated.
We recommend that the center coverage reference mark be on the centerline of the weld.
Содержание Reddy
Страница 1: ...User s Guide...
Страница 9: ...ix...
Страница 10: ...x www eddyfi com General Precautions and Conventions General Precautions and Conventions...
Страница 15: ...xv General Precautions and Conventions...
Страница 16: ...2 www eddyfi com Reddy Overview Chapter 1 Reddy Overview...
Страница 27: ...13 Reddy Overview...
Страница 28: ...14 www eddyfi com MagnifiGO Overview Chapter 2 MagnifiGO Overview...
Страница 34: ...20 www eddyfi com MagnifiGO Overview Figure 2 6 Acquisition Summary Figure 2 7 Report Summary...
Страница 35: ...21 MagnifiGO Overview Figure 2 8 Save Figure 2 9 Material Database...
Страница 36: ...22 www eddyfi com Preferences Chapter 3 Preferences...
Страница 38: ...24 www eddyfi com Preferences Figure 3 3 Analysis preferences Figure 3 4 Help...
Страница 39: ...25 Preferences...
Страница 40: ...26 www eddyfi com Keypad and Keyboard Functions Chapter 4 Keypad and Keyboard Functions...
Страница 42: ...28 www eddyfi com Using Reddy with Sharck Probes Chapter 5 Using Reddy with Sharck Probes...
Страница 62: ...48 www eddyfi com Maintenance and Troubleshooting Chapter 6 Maintenance and Troubleshooting...
Страница 66: ...52 www eddyfi com Specifications Chapter 7 Specifications...
Страница 69: ...55 Specifications...
Страница 70: ...56 www eddyfi com Connector Reference Appendix A Connector Reference...
Страница 74: ...60 www eddyfi com Using the Optional Harness Appendix B Using the Optional Harness...
Страница 82: ...www eddyfi com...