Chapter 4 — The QuadStat
45
is not for grounding the QuadStat itself. If your Faraday cage is already
earthed by its own ground connection then you should not use the
QuadStat ground cable! Use of the QuadStat cable in this instance will
provide a second pathway to earth which could result in a ‘ground
loop’ which can actually increase signal interference!
To e-corder Inputs”
2 1
To
e-corder
Input:
8
7 6
5 4
3 2
1
To
e-corder
Input: 8 7
6 5
4 3
2 1
For single channel operation use only Channel 1.
For bipotentiostat operation use only Channel 1 and Channel 2.
To e-corder Output or
waveform generator
Single channel operation
Multiple channel operation with the same
applied waveform on each channel.
To e-corder Output or
waveform generator
Figure 4–6
Using the external inputs
of the QuadStat. Note
connection to E In is only
required for potentials of
more than ±2.5 V, or for
pulsed or ramped
waveforms.
Figure 4–5
Signal connections from
the terminal adaptor.
Black colored wires are
connected to COM pins.