VE-H100B Veterinary Pulse Oximeter User Manual
- 60 -
5.2.3 Alarm for SpO
2
Sensor off
When the SpO
2
sensor falls off from the finger, the oximeter will
give a medium alarm, and the icon
displays in
information area.
SpO
2
and PR value area display “---”, and give a medium alarm.
(Make sure the alarm system in menu is ON.)
5.2.4 Alarm for Low Battery
When the battery is too low, the oximeter gives a medium alarm
for low battery.
After the low battery alarm occurs, the oximeter can still be
operated for a few minutes before it turns off automatically.
The low battery icon
displays on LCD, and gives a
medium alarm. (Make sure
Alarm System
in menu is ON.)
5.2.5 Higher than Hi Alarm Limit
If the measured SpO
2
or PR value is higher than the Hi Alarm
(upper alarm limit), the oximeter gives a high alarm.
Here we take PR for example:
If the measured PR value is higher than the setup
PR Hi Alarm
,
the oximeter gives a high alarm (Make sure
Alarm System
in
menu is
ON
). A ↑ icon displays near PR, which indicates that the
measured value is higher than that of
PR Hi Alarm
, it will