VE-H100B Veterinary Pulse Oximeter User Manual
- 59 -
Alarm Indicators
When an alarm occurs, the oximeter will indicate it through the
following indications:
■
Character flash
■
Alarm tone
High level alarms: character flashes quickly and sounds triple +
double beep;
Medium level alarms: character flashes slowly and sounds triple
beep;
The sound pressure range for auditory alarm signal is from 45 dB
to 85 dB.
5.2 Alarm Conditions
5.2.1 Alarm off Before the First Measurement
Before the first measurement, the alarm system is configured to be
off. At this time, if the SpO
2
sensor is unconnected or the sensor is
off, the oximeter will not give an alarm.
5.2.2 Alarm for SpO
2
Sensor Unconnected
When the SpO
2
sensor is disconnected, the oximeter gives a
Medium alarm. The icon
displays in information area.
SpO
2
and PR value area display “---”, and give a medium alarm.
(Make sure the alarm system in menu is ON.)