![EDAN F15 Series Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/edan/f15-series/f15-series_service-manual_2357161065.webp)
F15 Series Fetal & Maternal Monitor Service Manual
Disassembling the Monitor
- 54 -
9.4.6 Replacing the iM X12 Parameter Board (wired model only)
To remove the iM X12 parameter board,
1)
Separate the upper assembly and the lower assembly.
2)
Disconnect all the wires.
3)
Remove the four screws on the iM X12 parameter board.
To replace the iM X12 parameter board,
Replace the iM X12 parameter board in a reverse procedure. Refer to the following illustration when
connecting wires. Ensure they are all correctly connected.
1
2
3
4
5
6
7
8 9 10 11
No. Connect to
1 alarm indicator/indicator board
2 touchscreen
3 screen
4 printer head
5 out-of paper checking board
6 motor
7 paper drawer in place checking switch
8 maternal parameter board
9
loudspeaker
10 bottom interface board
11 top interface board