Operation
Page 91
3.4.34
Hardware Menu
Advanced Menu
→
Hardware Menu
This menu contains factory hardware installations. If the user reset to factory defaults, then they may
need to revisit this menu to enable their installed optional features.
Figure 76
–
Hardware Menu Screen
Model
Select the instrument model. Normally this only needs to be reset
when the configuration is corrupted. The selections available will
depend on licensing. It is not recommended to run an instrument
with firmware set to an incorrect model.
Front Panel Style
Choosing the incorrect front panel will result in the traffic lights
behaving inconsistently. Default is
Aluminium
.
Network Port
When
Enabled
indicates the instrument has a network port
installed. Default is
Disabled
.
Orifice Size
[Internal Pump Option Disabled]
Specify the input orifice for instruments without an internal pump.
Default is
1.08
for
Standard
range instruments and
1.50
for
High
range instruments.
Internal Pump
When
Enabled
indicates the instrument has an Internal pump
installed. Default is
Disabled
.
Input 1 is CO2
When
Enabled
, analog input 1 is treated as a measurement of CO
2
.
This gas will appear on the
Home screen
and can be logged as the
second gas. Default is
Disabled
.
External Bkgnds
When
Enabled
the system will not perform automatic backgrounds
and it will not warn the user that it is not doing backgrounds. Rather
it is the responsibility of the operator to remotely trigger the
backgrounds. Default is
Disabled
.
IRSource Setpoint
Setpoint target for the IR Source voltage. At power up, the Source
Pot will slowly adjust until this setpoint is reached. This is normally
set by the factory.
Cooler Setpoint
Setpoint target for the Cooler Voltage in the IR Detector. At power
up, the Cooler Pot will slowly adjust until this setpoint is reached.
This is normally set by the factory
Содержание Acoem Serinus 30
Страница 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Страница 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Страница 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Страница 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Страница 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Страница 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Страница 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Страница 162: ...Serinus 30 User Manual 3 3 Page 162 7 1 Flow Fault Figure 142 Flow Fault Diagnostic Procedure Section 2 3 2 Section 5 2...
Страница 165: ...Troubleshooting Page 165 7 4 Mirror Temperature Failure Figure 145 Mirror Temperature Failure Diagnostic Procedure...
Страница 166: ...Serinus 30 User Manual 3 3 Page 166 7 5 USB Memory Stick Failure Figure 146 USB Memory Stick Failure...
Страница 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Страница 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Страница 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Страница 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Страница 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Страница 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Страница 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Страница 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Страница 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Страница 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Страница 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...