Introduction
Page 21
1.
Introduction
1.1
Description
The Ecotech Serinus 30 Carbon Monoxide Analyser uses Non-Dispersive Infrared Spectrophotometry
(NDIR) technology to measure CO in ambient air, in the range of 0 - 200 ppm.
The U.S. EPA has designated the Serinus 30 Carbon Monoxide Analyser as a reference method and TÜV
has designated it as an EN approved instrument.
This section will describe the specifications of the instrument as well as the main components and
techniques used to obtain stable gas concentration readings.
1.2
Specifications
1.2.1
Measurement
Range
0 - 200 ppm (autoranging)
U.S. EPA designated range:
Any full scale range between 0 - 50 ppm
TÜV EN certification range:
0 - 100 ppm
Lower Detectable Limit
< 0.04 ppm (40 ppb), with Kalman filter active
1.2.2
Precision/Accuracy
Precision
20 ppb or 0.1% of reading, whichever is greater
Linearity
Better than ± 1% of full scale (0 - 50 ppm); ± 2% of full scale (0 - 200 ppm), from best straight-line fit
Noise at Zero
< 0.02 ppm or 0.2% of concentration reading whichever is greater
Response Time
60 seconds to 95%
Sample Flow Rate
1.0 slpm
Содержание Acoem Serinus 30
Страница 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Страница 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Страница 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Страница 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Страница 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Страница 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Страница 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Страница 162: ...Serinus 30 User Manual 3 3 Page 162 7 1 Flow Fault Figure 142 Flow Fault Diagnostic Procedure Section 2 3 2 Section 5 2...
Страница 165: ...Troubleshooting Page 165 7 4 Mirror Temperature Failure Figure 145 Mirror Temperature Failure Diagnostic Procedure...
Страница 166: ...Serinus 30 User Manual 3 3 Page 166 7 5 USB Memory Stick Failure Figure 146 USB Memory Stick Failure...
Страница 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Страница 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Страница 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Страница 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Страница 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Страница 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Страница 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Страница 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Страница 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Страница 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Страница 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...