24
During use, always make sure that the leather pouch is completely closed and
undamaged.
- Only accessories approved by ecom instruments GmbH may be used.
- Avoid using the unit in aggressive acidic or alkaline solutions.
- Ensure that the unit is not taken into Zone 0 areas.
4.1 Special conditions for use in Zone 0 areas
Temperature measurement with the Ex-MX2 is only allowed within Zone 0 only if the
following conditions are met:
1. The Ex-MX2 temperature measuring device itself must not be brought into Zone 0.
2. Absolutely ensure that the device cannot fall into Zone 0, even by mistake. To ensu-
re this, the device’s carrying sling must also be attached securely to the operator’s
belt or to a belt loop.
The length of the carrying sling should be measured so that the Ex-MX2 still
maintains a distance of 50 cm to the boundary area of Zone 0 when it hangs freely
on the carrying sling.
Only the accessories approved by ecom instruments GmbH may be used.
5. Safety instructions
- Caution: laser beam!
- Do not look directly into the laser beam!
- Do not aim the beam at other people!
- Danger of eye damage!
- Laser class 2
- Power:
< 1 mW
- Wave length: 635 nm
6. Ex data
EC-type examination certificate no.: TÜV 00 ATEX 1597 X
Ex designation: II 2 G EEx ia IIC T4
Approved for Zones 1 and 2, device group II, gas group C potentially explosive gases,
vapours or mist, temperature class T4.
7. Technical data
Measuring range:
-30 ... +900 °C
Ambient temperature:
0 °C ... +50 °C; laser operation up to m45 °C
Storage temperature: -20 °C ... +50 °C (without batteries)
Batteries / power supply:
2 x LR06 (AA) according to IEC
Emissivity:
0.10 ... 1.50 (0.95 preset)
Measuring spot marking:
3 point laser marking
Содержание Ex-MX2
Страница 1: ...Safety instructions Ex MX2 ...
Страница 15: ...35 16 EC Type Examination Certificate 19 ...
Страница 16: ...36 ...
Страница 17: ...37 ...
Страница 18: ...38 ...
Страница 19: ...39 ...
Страница 20: ...40 ...