8. Flue Gas Analysis
8.1. Gas analysis
After the 1-minute calibration phase, the instrument switches over to
the measurement mode. The gas measurement values can be
viewed on 4 display pages. Use the cursor keys to scroll the pages.
Pressing <F1> enables to switch, from the values display, to a menu
point selected before or on „Standby“ (see chapter Adjustments).
Possible menu points are: Soot...Oil trace, Data processing, View
memory, Display values, Fuel type, Efficiency (K), Internal, Adjust-
ments. Furthermore you can switch back from any random menu
point to the values display with <F1>.
Press <F2> to print out the values simultaneously to their recording
in the intermediate memory.
Press <F3> to cut the CO sensor from the gas path so that it is pro-
tected against too high concentrations.
Pressing <F4> also enables to switch, from the values display, to a
menu point selected before or on „Standby“ (see chapter Adjust-
ments). Possible menu points are: Soot...Oil trace, Data processing,
View memory, Display values, Fuel type, Efficiency (K), Internal,
Adjustments (back to values display with <F1>).
Hotkey
Key <F1>
Re-calibrate the instrument after each
measurement (after one hour at the latest)
to keep the accuracy of the results!
Hotkey
Taste <F4>
Store and print
values
key <F2>
Switch-off CO
sensor
key <F3>
Symbol
cooler
ON/OFF
O2
3.2
%
CO2
13.1
%
T.Gas
184
°C
T.Air
20
°C
BImSchV