ECP2050
57 / 64
9. R
EPLACEMENT OF
L
INE
F
USE
Disconnect the unit from power
source!!!
Using a flat-tip screwdriver release
the fuse holder from its bottom.
Take out the fuse holder.
Replace the old line fuse.
In the holder insert new fuse from
unit accessories,
its value is
stated on unit label.
Insert the fuse holder back.
Push the fuse holder all the way
inside.
10. T
ROUBLESHOOTING
Problem
Caused by / Conditions
Solution
Leaking liquid from washing
heads.
Damaged seal in pumping
head.
Replace the seal
immediately. The pump must
not be used with damaged
seal.
Pressure pulsation.
Pressure sometime suddenly
decreases.
Column, injection valve or
other equipment connected
after the pump.
Test if the problem persists
without any equipment after
the pump. Then try to connect
any reliable source of
hydrodynamic resistance
(column, reduction valve).