ECP2010
/
ECP2010H
46 / 68
6.5.6.1. Seals
L.PUMPING
- left working head seal,
R.PUMPING
- right auxiliary head seal,
L.WASHING
- left washing head seal,
R.WASHING
-right washing head seal.
6.5.6.2. Check Valves
INLET
- inlet check valve,
OUTLET
- outlet check valve.
6.5.6.3. Pistons
LEFT
- left piston,
RIGHT -
right piston.
6.6. Diagnostics
In each submenu is displayed diagnostic information about the unit.
6.6.1. Errors
On screen is displayed
ERROR
. Unit knows three types of errors
–
simple error, operating
error and fatal error. When reporting simple error, pumping does not stop, error is announced by
three beeps and LED ERR still shines red. When reporting operating eroor, pumping stops and
after pressing PUMP it starts running again, if the fault source has been corrected. When reporting
fatal error, pumping stops and its start is blocked. Reset can be done by turning the unit off and on,
after elimination of the fatal error.
If an error occurs, display shows symbol
E
with a
number of errors.
When entering
Menu/Diagnostics/Errors
a
number is displayed (
ERROR NO:
) and a
description of actually displayed error, total
number of errors (
TOTAL:
) and eventually arrows
suggesting possibility of listing, when there is
more then one error. When entering this detailed
screen, an actual status is displayed and it is not
updated. For updating this information, it is
necessary to re-enter this screen.
For quick access to error and warning
descriptions, the
QUICK DIAGNOSTICS
screen
is displayed, which can be called from the main
screen by pressing
ENTER
. When entering this
detailed screen, an actual status is displayed and
it is not updated. For updating this information, it
is necessary to re-enter this screen.
6.6.2. Warnings
Screen displays
WARN
.