Ecom-D
Page 31
After sufficient purging time (X behinds CO disappears), the sensor
can be reintegrated into the measurement system with
<F3>
(if you
choose
<Yes>
in
<Adjustments> / <Internal> / <CO-Automatic>,
the CO sensor will begin measuring again automatically). The sensor
can also be manually excluded from the measurement system with
<F3>
.
Probe and tubing
Depending on the frequency of use, the probe and sample line should
be regularly cleaned in order to prevent particle deposit build-up and
to prevent early wearing due to corrosion.
-Release the connections at the instrument and at the probe grip
to disconnect the tubing.
-Clean it with warm water in then blow dry the water out
-Slightly grease the O-rings of the tubing connections from time to
time with acid-free grease.
Change printer paper roll
-Release the printer cover (press lock downwards).
-If necessary, extract the remaining paper from the printer. To do this,
select
<Adjustments>
/
<Paper feed>/<Enter>
).
-Remove the plastic tube of the previous roll.
-Insert the paper end in the slot under the transport roll (paper
roll on the inner side facing you when inserting the paper).
-Feed approx. 3 cm paper through the printer (
<Adjust-
ments>
/
<PaperFeed>
/
<Enter>
).
-Lay the paper roll in the corresponding space.
-Insert the paper through the slot of the printer compartment then cov-
er and close the lid.
Unlocking
printer cover