MANUAL DE USUARIO
– PROCAM HD II
Página
21
de
44
INMANUUML043 USER M.M. USO PROCAM HD II V.2023.07 ENGESP US
Image is in
black
and
white
Camera
is
defective.
Please contact to the nearest Technical Service.
Image without
real colors
Incorrect White
Balance.
Adjust White Balance.
Blurred Image
Lack of focus
adjustment of the
external
optical
system.
Perform the focus adjustment of the optical
system.
Check if the videocamera has the adapter ring.
FIBERCOSPE
function
activated.
If you don’t use flexible scope, deactivate
FIBERSCOPE function.
Pressing FIBERCOSPE button
Image is too
bright.
Misadjusted
monitor.
Calibrate properly the image levels of the monitor
Misadjusted
function.
Select the adequate specialty by pressing the
button
Image is too
dark.
Not enough light.
Increase the levels of intensity of illumination of the
light source.
Verify that the external optical system is clean.
Verify that the video monitor is configured with
adequate levels of brigthtness.
Increase the level of brightness by pressing the
button
If the inconvenience cannot be solved, please contact your local distributor or the nearest
ECLERIS office and ask for technical service.
9. TECHNICAL SERVICE
The device should be repaired and/or maintained exclusively by ECLERIS trained and qualified
personnel in authorized installations and must only be repaired the parts of the device that
ECLERIS state as repairable. ECLERIS will provide under request circuits, list of components,
description calibration instructions or any other information in order to assist the ECLERIS
authorized technical personnel.
We recommend using the following procedure to expedite repair service and reposition of
products under guarantee.
1. Contact the technical service department of your local distributor or nearest ECLERIS
office.