7. TROUBLESHOOTING
The following chart indicates the most frequent problems and their quick solutions. Please, in
case of any inconvenience, consult this chart.
PROBLEM
PROBABLE CAUSE
ACTION
The equipment does
not turn on
It is not connected to the
mains
Connect the power cable to
the mains
The power cord is broken or
cut.
Replace the power cord.
The plug is not in good
condition.
Change the plug.
The fuse is broken or
burned.
Replace the fuse.
Power switch Off.
Turn on the switch.
The green pilot light of
the ignition switch is
turned on but the
display does not turn
on.
System “hung”
Turn the unit off and on.
System with general failure
Contact the nearest Technical
Service.
The display works
normally but does not
respond correctly to
the vacuum circuit.
Vacuum pump does not turn
on
Turn the equipment off and
on, if the problem is not
solved, contact the nearest
Technical Service.
Vacuum pump ignites but
the vacuum circuit has leaks.
If the vacuum level is
insufficient there may be a
leak in the outer hoses and /
or connections of the vacuum
trap, check the connections.
Clogged vacuum trap cap
Verify that the rubber disk of
the vacuum trap is not
clogged sealing the vacuum
inlet of the trap.
Infusion circuit does
not infuse product
Infusion circuit stopped.
Press the pedal once to
activate the infusion circuit.
The infusion pump is not
working well on the hose.
Adjust the position of the
silicone hose on the peristaltic
pump and check the pump
hose locks for proper
adjustment.