USER MANUAL
–
Page
10
of
18
INMANUUIN155 Rev.2019.02.14 / ENG
7. With the button
the vacuum circuit is switched on and stopped, independently.
8. Remember that when the equipment is turned on, only the suction circuit is activated. To
activate the infusion circuit, use the pedal. When pressing the pedal once, the infusion
circuit is turned on, pressing again stops the infusion circuit. The operation of the pedal is
of the ON / OFF type, it must not be kept pressed. When the product infusion circuit is
active, the front panel lights blink. If the equipment is not infusing, the lights remain on
permanently.
The operation of the pedal is of the ON / OFF type, never keep it pressed.
9. To end the treatment, press the pedal to stop the infusion circuit. Next, stop the vacuum
circuit by pressing the button
.
10. If the composition of the product to be infused changes, a cleaning cycle must be carried
out.
11. If no further treatments are carried out, a cleaning cycle must be carried out.
5. OPERATION
– FUNCTIONS
5.1.
CLEANING CYCLE
It is very important to carry out cleaning cycles at the end of the day, since the residues of
products to be applied to the skin, epidermal cells and other treatment products could block
the lumen of the vacuum and / or infusion hoses.
The cleaning cycle is a process that cannot be interrupted by the user once it starts.