background image

 

USER MANUAL

 

EVERLUX

 

LED LIGHT SOURCE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ABOUT THIS MANUAL 

Versiion: 2019.04.01 

Year: 2019 

Language: English 

All the rights to this manual are reserved, especially reproduction, publication and translation rights. No portion of this manual may be 
reproduced by no means (photocopy, microfilm or other methods) without previous written authorization from ECLERIS. Neither can it 
be reproduced, multiplied or published through any electronic data processing form. 

The  information  contained  in  this  manual  may  be  modified  or  broaden  without  previous  notification,  without  it  representing  an y 
responsibility to ECLERIS. 

 

 

EUSA GLOBAL Llc. 

11801 NW 100th Rd 
Suite 17 
Medley, FL 33178, U.S.A. 
Phone: +1 786 483 7490

 

 

CMC Medical Devices & Drugs S.L. 

C/  Horacio  Lengo  Nº18,  CP29006,  Málaga, 
Spain 
Ph. +34951214054 
Email: [email protected]

 

 

 

 

 

 

Содержание EVERLUX

Страница 1: ...ut previous written authorization from ECLERIS Neither can it be reproduced multiplied or published through any electronic data processing form The information contained in this manual may be modified...

Страница 2: ...3 2 INTENDED USER 5 3 3 FEATURES AND DESCRIPTION 5 3 3 1 FRONT PANEL 5 3 3 2 REAR PANEL 5 3 4 COMPONENTS 6 4 INSTALLATION AND SET UP 6 4 1 REVISION 7 5 OPERATION FUNCTIONS 7 6 MAINTENANCE 7 6 1 CLEAN...

Страница 3: ...recautions Please read the manual thoroughly before operating the equipment Should you have any doubt regarding the contents of the manual herein or the equipment operation and safety please contact y...

Страница 4: ...blem the system must be laid a plane surface and power cable disconnected The device must be used only by personnel informed of all this warnings Any use outside of the manufacturer specific intent us...

Страница 5: ...its use for all types of procedures The EVERLUX allows the connection of the Ecleris PROCAM universal endocamera which has a 1 3 CCD sensor and integrated controls especially designed for medical appl...

Страница 6: ...l match perfectly and not exceed the adapter 5 In case of using the EVERLUX as a Microscope Colposcope light source mount the EVERLUX on the Ecleris Microscope Colposcope forearm See Light Source Inst...

Страница 7: ...S office prior to its use 5 OPERATION FUNCTIONS This section thoroughly describes the device functioning This system must be used only for medical paramedical or health professionals 1 Connect the pow...

Страница 8: ...following the instruction provided by their manufacturer Do not turn the light source on until an hour 1 has passed from the cleaning since the unit can get damaged It is recommended to clean the exte...

Страница 9: ...UT Composite video CVBS 1 Vpp 75 BNC S Video Y C Jack mini DIN ECLERIS ENDODIGI CONNECTOR Jack 3 5mm mono DIMENSIONS 120x300x155mm width x depth x high MASS 7 7 pounds SERVICE CONDITIONS 50 104 F 35 7...

Страница 10: ...Lateral ventilation grids are obstructed Clear the lateral ventilation grids Fan is not working Turn the equipment off immediately and contact the closest Technical Service Light is correct but no im...

Страница 11: ...blem description The shipment should be done in the original packaging to avoid damage to the equipment during the transportation ECLERIS will not be responsible for damage caused during the transport...

Отзывы: