30
3.4.5
SENSOR
N
AME
Two 16 character lines are available for naming the Sensor, Upper and Lower case characters, Numbers and
Punctuation are available. The information entered will be displayed in the INFO screen. Entry is accomplished
by scrolling through the alphanumeric list with the
▲ (
forward →) and
▼ (
backwards ←) arrows to the
character of choice and then pressing NEXT to advance the cursor to the next digit. Pressing and holding the
▲
or
▼
keys will initiate two speed auto scrolling. Press BACK to exit the screen.
Location: CONFIG → XMTR → LCD → LABELS → SENSOR
3.4.6
P
ASSWORD
P
ROTECTION
PASSWD Enter 4 character password to protect access to MENU Level, CAL Menu, CONFIG Menu and SIM Menu
(simulate). Each level can be turned ON or OFF and can have a unique password. Upper Case Characters and
Numbers are available for use.
Place the cursor in front of the level to be changed and Press OK. Move the cursor to
ON and press OK to change the password status from OFF to ON.
Entry is accomplished by scrolling through the alphanumeric list with the
▲ (
forward
→) and
▼ (
backwards ←) arrows to the character of choice and then pressing NEXT
to advance the cursor to the next digit. Pressing and holding the
▲
or
▼
keys will
initiate two speed auto scrolling.
o
MENU
ON/OFF
_ _ _ _
Locks Main Menu
o
CAL
ON/OFF
_ _ _ _
Locks CAL and
CONFIG
o
CONFIG
ON/OFF
_ _ _ _
Locks CONFIG
o
SIM
ON/OFF
_ _ _ _
Locks SIM and CONFIG
In the case of a Lost or Forgotten password enter MSTR to access the screen.
Location: CONFIG → XMTR → PSSWD