OPERATING INSTRUCTIONS
MODEL: 111
INFRARED THERMOMETER
WITH LASER SIGHTING
INTRODUCTION
This instrument is a portable easy to use 3½digit, com-
pact-sized digital infrared thermometer with laser sighting
designed for simple one hand operation. Meter comes with
backlight LCD display, Auto-Hold function and Auto
Power Off (10 seconds approx.) to extend battery life.
SAFETY INFORMATION
It is recommended that users read the safety and opera-
tion instructions before using the infrared thermometer.
DANGER
When
sign appears, pull the trigger to turn the laser
beam on and off.
Be very cautious with the laser beam, not aiming your
eyes or those of any other person or animal.
●
Do not look directly into the laser light from the opti-
cal system.
●
When measuring the temperature of an object which
has a mirror finish, be careful not to allow the laser
light beam to be reflected off the surface into your
eyes or those of another person.
●
Do not allow the laser light beam to impinge upon any
gas which can explode.
CAUTION
●
Do not use the unit near any device which generates
strong electromagnetic radiation or near a static elec-
trical charge, as these may cause errors.
●
Do not use the unit where it may be exposed to corro-
sive or explosive gases. The unit may be damaged, or
explosion may occur.
●
Do not keep or use this unit in an environment where
it will be directly illuminated by sunshine, or where it
condensation. If you do, it may be deformed, its insu-
lation may be damaged, or it may no longer function
according to specification.
●
Do not place the meter on or around hot objects
(70°C/158°F). It may cause damage to the case.
●
If the meter is exposed to significant changes in ambi-
ent temperature (hot to cold or cold to hot). Allow 30
minutes for temperature stabilization, before taking
measurement.
●
Condensation may form on the sensor when going
from a cold to hot environment-wait 10 minutes for
condensation to dissipate before taking measurements.
●
This unit is not constructed to be waterproof or dust
proof, so do not use it in a very dusty environment or
in one where it will get wet.
●
Do not point the lens at the sun or at any other
source of strong light. If you do, the sensor may be
damaged.
●
Do not contact the lens against the object whose
temperature is to be measured, or get it dirty, allow
it to be scratched, or allow any foreign material to
adhere to it. Doing so may cause errors.
●
Do not touch or hold by the front case. Tempera-
ture reading can be affected by heat from hand.
●
Readings may be affected if the unit is operated
within a radio frequency electromagnetic field
strength of approximately 9 volts per meter, but
the performance of the instrument will not be per-
manently affected.
SPECIFICATIONS
GENERAL
Display:
3½ digit liquid crystal display (LCD) with maxi-
mum reading of 1999.
Low battery indication:
The “ ” is displayed when the
battery voltage drops below the operating level.
Measurement rate:
0.25 second per measurement.
Operating Environment:
32°F to 122°F (0°C to 50°C) at
< 70% R.H.
Storage Temperature:
-4°F to 140°F (-20°C to 60°C), 0
to 80% R.H. with battery removed from meter.
Auto power off:
10 seconds.
Standby consuming current:
<5µA.
Battery:
Standard 9V battery (NEDA 1604, IEC 6F22
006P).
Battery Life:
9 hours (continuity) typical.
(contain Laser and Backlit)
Dimensions:
148mm(H) x 105mm(W) x 42mm(D).
Weight:
approx. 157g. (including battery)
Laser Specifications
Laser safety classification of Class 2.
Wave Length:
Red (630~670nm).
Power out:
<1mW, class 2 laser product.
ELECTRICAL
Temperature Range:
-30°C to 550°C / -22°F to 1022°F.
Display Resolution:
0.5/1°C (Auto), 1°F.
Accuracy:
±(2°C/4°F) for -30°C to 100°C, -22°F to 212°F.
±(2% reading) for 101°C to 550°C, 213°F to 1022°F.
Temperature Coefficient:
±0.2% of reading or ±0.36°F/0.2°C, whichever is
greater, change in accuracy per °F/°C change in ambient
operating temperature above 82.4°F/28°C or below
64.4°F/18°C.
Response Time:
0.25 second
Spectral Response:
6 to 14µm
Adjustable emissivity (
ε
):
0.1 to 1.0.
Detection Element:
Thermopile.
Optical Lens:
Fresnel Lens.
Sighting:
1-beam laser marker <1mW (class 2).
Field of View:
100mmØ at 1000mm (3.9"Ø at 39.0").
Spot size increases with distance from the probe tip as shown
(Spot Diameter measured at 90 % Energy)
DEFINITION
FUNCTION KEY
SET key
Press “SET” key goes to the setting mode. Use the same
key to go to next step, following as the cycle :
ε →
ALM
Hi
→
ALM Lo
→
MAX
→
MIN
→
measuring mode.
key
Press key, LCD shows “ ” sign. Then pull the trig-
ger to turn on laser pointing function to aim at the measur-
ing point of object. Repeat same steps turns off laser point-
ing function.
key
Use key to turn on or off the Back-Light function.
°C/°F key
Use the °C/°F key to select reading displayed in Cel-
sius(°C) or Fahrenheit(°F).
APO key
The meter is default to APO (auto power off) activated
along with manual mode. Leave the meter without opera-
tion for 10 seconds, power turns off automatically.
Press “APO” key to disable APO function and will go to
auto continuous mode. Continuous measuring will be
performed without holding the trigger. Press “APO” key
again to return to APO activated along with manual mode.
▲
and
▼
key
Use
▲
or
▼
to adjust
ε
, ALM Hi, and ALM Lo in
setting mode. Hold the key will speed up.
Customer Care Australia P: 13122 F: 1300 303 310 E: [email protected] W:cabac.com.au
A: Distribution Pl, Seven Hills NSW 2147
Customer Care New Zealand P: 0800 966 969 F: 0088 6969 E: [email protected] W: cabac.co.nz