TTX 200 Type T Core Thermometer Operating Instructions 

Congratulations on the purchase of the product. We hope that you can make use of this product for a long time and it 
helps you with your work. Powered by a replaceable lithium battery, the thermometer is a handy, temperature 
measurement device for a whole range of measurement and control applications. 
 
 

  LCD 

  ON/HOLD 

  Thermocouple Probe

 

  Battery Cover 

 

OPERATION 

 

1. Power on 

To switch the thermometer on,   

  keep ON/HOLD (

) pressed   

for approximately one second. 
 

2. Display LCD 

All necessary information is   
displayed  on  the  LCD.  The  individual  symbols  have 
the following meanings: 
2.1 Display for current   

measured values   

2.2 Minus sign 

        (negative measured values) 

2.3 Battery status indicator 

  2.4 

HOLD

 -- Auto shut down after 10 minutes 

2.5 Temperature unit 

 

3. Select the required unit of measurement °C or °F:

 

Switch  on  the  device  by  pressing  ON/HOLD  (

).  As  soon  "Ebro"  is  shown,  press  ON/HOLD  (

)  again  for  3     

seconds and release. The settings menu is active and the current unit is shown. 

Press ON/HOLD (

) once, the display will switch to next unit. 

Press ON/HOLD (

) for more than 2 sec, the current unit will be stored, and then the settings menu will be left. 

 

4. Measuring temperature   
  4.1 As soon as switch on the device, start to take a temperature reading every second.   
  4.2 Press ON/HOLD (

) once, the device stops measuring and the last reading is shown on the LCD.     

       

“HOLD” is shown on the LCD. 

 

  4.3 Press ON/HOLD (

) again, the device starts measuring again. 

* The last measured value is frozen and the thermometer deactivates itself automatically after about 10 minutes to   

save battery power. 

 

 

CAUTION 

The thermometer should be protected from the following:   
1. Electro Static Discharge   
2. 

”Thermal shock” caused by large or abrupt

 ambient temperature changes. 

- allow 30 minutes for unit to stabilize before use w

hen exposed to “thermal shock”.

 

3. Do not leave the unit on or near objects of high temperature. 
4

Keep out of reach of all children.

 

 

STORAGE & CLEANING 

Clean the device with a damp cloth. Do not use any solvents such as Acetone as they corrode the plastic. Isopropyl alcohol may be 
used  to  disinfect.  Do  not  submerge  any  part  of  the  thermometer.  The  thermometer  should  be  stored  at  room  temperature 
between 

20 to +65°C (-4~149°F). 

 

 
LCD ERROR MESSAGES 

The thermometer incorporates visual diagnostic messages as follows: 

ˬ

˭

 

ˬ

˭

 

‘Er2’ is displayed when the thermometer is exposed to rapid changes in the ambient temperature. 

The 

thermometer

 

should  be  allowed  plenty  of  time  (minimum  30  minutes)  to  stabilize  to  the  working/room 

temperature. 

 

‘Er3’ is displayed when the ambient

 temperature exceeds -20

o

C(-4°F) or +60

o

C(140°F).   

ˬ

˭

 

Error  5~9,  for  all  other  error  messages  it  is  necessary  to  reset  the  thermometer.  To  reset  it,  turn  the 
instrument off, remove the battery and wait for a minimum of one minute, reinsert the battery and turn on. If 
the error message remains please contact the Service Department for further assistance. 

ˬ

 

˭

ˬ

˭

 

Hi

 or 

Lo

 is displayed when the temperature being measured is outside of the measurement range. 

 

BATTERy

 

The thermometer incorporates visual low battery indication as follows: 

 

 

 

‘Battery OK’: measurements

 

  are possible 

‘Battery Low’: battery needs to be 

 

replaced, measurements are still possible 

‘Battery Exhausted’: measurements are 

not possible 

BATTERY REPLACEMENT 

When the 

Low Battery

 icon indicates the battery is low, the battery should be replaced immediately with a CR2450 lithium cell. The 

battery is located under the twist cover at the rear of the thermometer. Please note: It is important to turn the instrument  off before 
replacing the battery otherwise the thermometer may malfunction. 

  Dispose of used battery properly and keep away from children. 

 

SPECIFICATION 

Measurement Range 

-30~199.9°C (-22~391.8°F) 

Operating Range 

-20~60°C (-4~140°F) 

Accuracy 

 

± 0.5°C(0.9°F) at -30~100°C (-22~212°F),   
1% of reading at 100.1~199.9°C (212.1~391.8°F) 

Resolution 

0.1°C/0.1°F   

Battery Life 

>500h at room temp. and >250h at -20°C with measurement active. 

Dimensions 

127 x60x16 mm (5×2.36×0.63 inch) 

Needle 

120mm length, 2mm diameter of tip, 3mm diameter of shaft 

Weight 

120.7 grams(4.26 oz) including a CR2450 lithium cell 

 

  Caution: Please make sure the target to be measured will not exceed the temperature range of the probe to avoid 

permanent damage of the thermocouple probe. 

  Caution: To avoid electric shock and thermometer damage, do not measure live circuit where voltage exceeding 24V   

AC RMS or 60V DC with the thermocouple probe. 

 

EMC/RF

I

: Readings may be affected if the unit is operated within radio frequency electromagnetic field strength of 

approximately 3 volts per meter, but the performance of the instrument will not be permanently affected.

 

 

Xylem Analytics Germany GmbH - ebro 
Dr.-Karl-Slevogt-Strasse 1 
D-82362 Weilheim 
Germany 

* IP65 Splash Proof 

 