Measurement current selection
The unit offers the possibility to select ground continuity test current (step of 0.5 A).
The test current value selection is done according to the standards.
Move the reverse video line in front of the CURRENT line.
Press on the RIGHT arrow or ENTER key.
Display shows :
Fig 9.3
Select the number to be modified by moving the cursor with the RIGHT/LEFT arrow keys.
Increase or decrease the value with respectively the UP and the DOWN arrow keys. Repeat
operation for all the numbers if necessary.
Enter the current with the ENTER key. If the value is higher than the maximum limits (5.0A-
32.0A), following error message appears:
LIMIT ERROR
.
Enter a correct value or [ESC]
Open circuit test voltage selection (6 or 12 VAC)
The unit offers the possibility to select ground continuity open circuit test voltage (6 or 12 VAC).
The test voltage selection is done according to the standards.
Move the reverse video line in front of the VOLTAGE line.
Press on the RIGHT arrow or ENTER key.
Display shows :
Fig 9.4
With the UP/DOWN arrow keys scroll all the available voltages (6, 12).
Enter the selected value with the ENTER key.