XS series operating manual
XS user manual v0.55
- 24 -
1000V megohmmeter function (option XS26 or MXS1000)
Measurement voltage
Voltage adjustable from
20V
to
1000VDC
by 1 volt step.
Accuracy :
±(1% + 1V).
Positive pole of the HV generator is grounded (earth)
Dynamic stability for a mains
ΔV =
±10% : <±1%.
The maximum current in the measurement circuit is :
2 mA ±20%.
The discharge current is limited by a
2.2 k
resistor.
Measurement range
Display :
2000 digits
The measurement range is defined by the following formula:
(Utest / Umax generator ) x 200 G
*
* or
2T
according to the
option
This gives the following board for the
100, 250, 500, 1000VDC
standard voltages
200G
basic version
2T
option version
Voltage
Measurement range
Voltage
Measurement range
100V
100k
to 20G
100V
100k
to 200G
250V
250k
to 50G
250V
250k
to 500G
500V
500k
to 100G
500V
500k
to 1T
1000V
1M
to 200G
1000V
to 2T
Measurement accuracy
2000 digits numeric display with unity indication
(K
, M
, G
,T
).
Accuracy (% of the reading, 1U = 1 digit ) :
DXS and SXS models
200 G
basic version
(1.5% + 1U)
2 T
option and U<=200VDC
(2% + 1U)
2 T
option and U > 200VDC
(1% x Utest / 100 + 1U)
CAPACITANCE mode : from
1.00M
to 200G
(or
2 T
for option
20
) with an accuracy equal to the
NORMAL mode accuracy
±100k
Input impedance =
10 M
±1%