Obtaining the Developer Usage Rate
Description:
This procedure provides you with the average volume of developer used to replenish a single sheet of
film processed.
Procedure:
[1] From the Walk-Up Menu, press the “GO TO SETUP” key.
[2] Enter the 4-digit access code.
[3] Press the “MORE” key.
[4] Press the “INFO” key.
[5] Press the “USAGE” key.
[6] Press the “DEV USAGE” key.
[7] Press the “USAGE RATE” key. The average volume of developer replenishment used per single sheet of film
processed will appear on the Display Screen.
To clear the developer rate information, do the steps below.
[8] Press the “CLEAR USAGE” key to clear both the total volume usage and the usage rate.