ADJUSTMENTS AND REPLACEMENTS
Additional Service Procedures
06DEC05
AR4366-1
Page
201 of 219
Doing a Calibration of the LASER
Important
For systems with software version
≥
4.1,
you must have a “Session ID” for access to “Service
Functions” and “Diagnostics”. See SERVICE BULLETIN 843.
1
At the main menu, click:
•
[Service Functions]
•
[Diagnostics]
•
[LASER POWER]
2
Record the values from the LABEL for the MINI FAST SCAN AY.
3
Enter the values for:
•
“Power High”
•
“Voltage High”
•
“Power Low”
•
“Voltage Low”
4
Click:
•
[CALIBRATE NEW LASER]
•
[Save New Backup]
5
Do the adjustment for the
.