![East Tester ET-2715 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/east-tester/et-2715/et-2715_user-manual_3438341021.webp)
18
6 Specifications
Specifications are based on a one year calibration cycle and apply from +18
℃
to
+28
℃
unless stated otherwise. All specifications assume a 10 minute warmup period.
6.1 DC Voltage Measurement
Range
Maximum
measurement range
Resolution
Accuracy (% of r Counts)
ETX-1815
ETX-2015
30V
0V
~
31V
0.001V
0.05+2
0.02+2
100mV
-15mV
~
80mV
0.001mV
0.05+20
0.02+20
80mV
~
125mV
0.01mV
0.05+2
0.02+2
-10
℃~
18
℃,
+28
℃~
55
℃
temperature coefficient,±0.005%FS/
℃
.
Input resistance:
>
1MΩ.
6.2 DC voltage Source
Range
Maximum output
span
Resolution
Accuracy (% of r Counts)
ETX-1815
ETX-2015
100mV
-15mV
~
99.999mV
0.001mV
0.05+20
0.02+20
100mV
~
125mV
0.01mV
0.05+2
0.02+2
10V
0
~
11V
0.001V
0.05+2
0.02+2
-10
℃~
18
℃,
+28
℃~
55
℃
temperature coefficient,±0.005%FS/
℃
.
Maximum load: 1mA or 1kΩ
(
It should be based on the lower load.
)