Premier Series Monitor
RP-1000/RP-1200
TECHNICAL REFERENCE
MANUAL
EAGLE MICROSYSTEMS, INC.
366 Circle of Progress
Pottstown, PA 19464
PHONE: 610-323-2250 FAX: 610-323-0114
www.eaglemicrosystems.com
Страница 1: ... 14 FIRMWARE VERSION 2 32 Premier Series Monitor RP 1000 RP 1200 TECHNICAL REFERENCE MANUAL EAGLE MICROSYSTEMS INC 366 Circle of Progress Pottstown PA 19464 PHONE 610 323 2250 FAX 610 323 0114 www eaglemicrosystems com ...
Страница 2: ...djust the value of the parameter displayed ALM Go to the alarm set point menu CAL Go to the calibration menu ACK Acknowledge the alarm condition s CLEAR Temporarily clear the alarm condition s x y Set point x indicates a low or high set point L or H and y indicates the relay number 1 2 or 3 ZRO Enter the cell input zero menu SPAN Enter the cell span input set menu PO1 Calibrate the PO1 4mA Set 4mA...
Страница 3: ...ph that is represents the analog output An optional residual probe can be displayed on the main screen Pressing the SET pushbutton prompts the operator to enter a password If the password entered is correct and before the unit displays the METER SETUP menu pH 7 00 SET VALUE LABEL PUSHBUTTON LEGEND UNITS if needed PH ORP FL NH3 BARGRAPH pH 7 00 RE 0 00 ppm SET VALUE LABEL PUSHBUTTON LEGEND UNITS PH...
Страница 4: ...s the ADJ pushbutton to set the choice to METER if not set Parameter TEMPERATURE Selects the unit of the temperature display as either degrees Celsius or Fahrenheit Parameter BUZZER Enables ON or disables OFF the audible alarm Parameters K1 K2 K3 Sets the operating mode of the selected relay K1 K4 ENERGIZED DE ENERGIZED OFF NEVER ALWAYS LOW INPUT SP INPUT SP LOW w ACK INPUT SP INPUT SP OR Acknowle...
Страница 5: ...H etc Other labels that are specific to the probe input type will change accordingly Parameter ORP FS FLR FS pH FS NH3 FS or TTL FS Sets the value that represents an analog output of 20mA Parameter ORP ZERO FLR FS pH ZERO NH3 ZERO or TTL ZERO Sets the value that represents an analog output of 4mA Parameter PROBE DAMP This parameter sets the probe damping filtering The settings are OFF 1s 2s 3s 5s ...
Страница 6: ...energized based on the set point s mode value No screen will be displayed indicating the alarm and the buzzer will not sound PASSWORD After one exits the CALIBRATION menu one is first prompted to change the stored password If one does not want to change the password press the NO pushbutton If the YES pushbutton is pressed the SET new password screen is entered and now one can set the new password ...
Страница 7: ...ect 0 01 instead Calibrate the lower point using P1 and the higher point using P2 PO1 PO2 OUTPUT CALIBRATION To calibrate the analog output for channel one or two PO1 or PO2 one must connect either a DMM mA input or other device that can read the PO1 PO2 current output The device should have at least 0 01mA precision NOTE PO2 is an optional board installed inside the PS 1000 enclosure 1 From the c...
Страница 8: ...NATE CALIBRATION NOTE PV1 PV2 share common negative TB3 3 5 if J1 and J2 both have G selected SRC 24V SGND INDUCTIVE LOAD OPTIONAL SNUBBER OPTIONAL CAN BE DC 100ft WARNING THE CONTACT CANNOT SUPPLY ANY POTENTIAL 9 1 TEMP CELL 4 20mA PV2 PV1 B A TB3 Rx Tx TB2 EAGLE MICROSYSTEMS INC CHANNEL ONE 4 20mA SOURCE CHANNEL TWO 4 20mA SOURCE AC1 LAMP1 LAMP2 LAMP3 L1 C1 R1 RELAYS COM K1 K2 K3 K4 K5 K6 5 6 7 ...
Страница 9: ... PNP SWITCH PS 1000S DSP BLUE BLACK BROWN NPN SWITCH BOARD REVISION REV 0644 AND ABOVE NPN OUTPUT PS 1000S DSP PROBE FLOW SWITCH WIRING BLUE BLACK BROWN Cl2 Probe VOLTAGE FLOW SIGNAL GROUND FLOW SWITCH IN0 IN1 INPUTS G V 5 6 3 4 1 2 TB10 VOLTAGE FLOW SIGNAL GROUND FLOW SWITCH IN0 IN1 INPUTS G V 5 6 3 4 1 2 TB10 ...
Страница 10: ...644 AND ABOVE NPN OUTPUT SRC 24V PS 1000S DSP PS 1000S CPU SHIELD IS NOT CONNECTED PROBE FLOW SWITCH WIRING BLACK BLUE BLACK BROWN 1x5 Cl2 Probe J2 9 1 TEMP CELL EAGLE MICROSYSTEMS INC 4 20mA PV2 PV1 B A TB3 VOLTAGE FLOW SIGNAL GROUND FLOW SWITCH IN0 IN1 INPUTS G V 5 6 3 4 1 2 TB10 VOLTAGE FLOW SIGNAL GROUND FLOW SWITCH IN0 IN1 INPUTS G V 5 6 3 4 1 2 TB10 ...
Страница 11: ...000 MOUNTS ON THE PS 1000S CPU PO1 TRACKS PV1 TRACKS PV3 PS 1000S w PSC3000 TOTAL Cl2 Probe PV3 FREE Cl2 Probe PV1 PV3 4 20mA Vsrc TO DISPLAY PS 1000S DSP PS 1000S CPU SRC 24V 9 1 TEMP CELL 4 20mA PV2 PV1 B A TB3 The 1000ohm RTD is optional Probe BNC Connector 1x5 BNC RTD 4 20mA LOAD 600ohms 4 20mA LOAD 600ohms ...
Страница 12: ...Notes ...
Страница 13: ...w ACK K2 POLARITY N O K3 OFF PSC probe type K3 POLARITY N O PSC probe range K4 K6 POLARITY N O Alarm settings ALM menu PV1 probe type K1 RES SP 1 LOW 0 00ppm PV1 probe range K2 RES SP 2 HIGH 0 00ppm K3 RES SP 3 OFF PASSWORD 0 0 0 Note Parameters that do not need to be set depending on the modes selected will not be available leave blank Default hardware K1 K2 K3 K6 relays RS232 single 4 20 mA outp...