background image

14

User Manual MOP301 Digital Moisture in Oil Immersion Probe

The register block for the configuration of the customisable Modbus map consists of the registers 6001 

(0x1770) to 6010 (0x1779). For the blockwise query of the measured values behind Modbus registers 3001 

(0xBB8) to 3020 (0xBCB), the firmware accesses this configuration area and thus gets the information 

which measured value/status registers are to be output. A maximum of 10 user-defined registers can be 

mapped. The table below shows an example:

Registers ...

... with these  

assigned measurands ...

... map to registers ...

... mirrored from  

source registers

Dec

Hex

Meas.

Unit

Type

Dec

Hex

Dec

Hex

Function code 0x10

Function code 0x03 / 0x04

6001

0x1770

Status

-

INT16

3001

0xBB8

602

0x259

6002

0x1771

T

°C

FLOAT32

3002

0xBB9

1003

0x3EA

°C

FLOAT32

3003

0xBBA

1004

0x3EB

6003

0x1772

T

°F

FLOAT32

3004

0xBBB

1005

0x3EC

°F

FLOAT32

3005

0xBBC

1006

0x3ED

6004

0x1773

T

°C

INT16

3006

0xBBD

4002

0xFA1

6005

0x1774

aw

-

FLOAT32

3007

0xBBE

1035

0x46E

-

FLOAT32

3008

0xBBF

1036

0x46F

6006

0x1775

x

ppm

FLOAT32

3009

0xBC0

1141

0x474

ppm

FLOAT32

3010

0xBC1

1142

0x475

6007

0x1776

Saturation

%

FLOAT32

3011

0xBC2

1137

0x470

%

FLOAT32

3012

0xBC3

1138

0x471

6008

0x1777

aw

-

INT16

3013

0xBC4

4068

0xFE3

6009

0x1778

x

ppm

INT16

3014

0xBC5

4071

0xFE6

6010

0x1779

Saturation

%

INT16

3015

0xBC6

4069

0xFE4

3016

0xBC7

65536

0xFFFF

3017

0xBC8

65536

0xFFFF

3018

0xBC9

65536

0xFFFF

3019

0xBCA

65536

0xFFFF

3020

0xBCB

65536

0xFFFF

Tab. 7  Custom Modbus map example

5.5  Device Status Indication

If a critical error occurs, all Modbus values are set to NaN (according to IEEE754 for data type FLOAT32) 

or to 0x8000 (INT16).  It is possible to read out all status and error information via Modbus register 602 

(0x259). Errors are displayed in bit-coded form. If an event is present, the corresponding bit is set to 1.
Measured values outside the measureing range are limited with the corresponding limit value.

Error Bits Description

Recommended action

Bit 0 Error: Hardware T-sensing

Return the faulty unit to E+E for service

Bit 1 Error: Short circuit of T sensing element

1. Clean sensing head acc. to cleaning instructions

2. Return the faulty unit to E+E for service

Bit 2 Error: Open loop of T sensing element

Return the faulty unit to E+E for service

Bit 3 Error: Short circuit of moisture sensing element

1. Clean sensing head acc. to cleaning instructions

2. Return the faulty unit to E+E for service

Bit 4 Error:  Open loop of moisture sensing element  

or heavy pollution

1. Clean sensing head acc. to cleaning instructions

2. Return the faulty unit to E+E for service

Bit 5 Warning:  Polluted moisture sensing element or  

highly conductive oil

Clean sensing head or oil is incompatible

Bit 6 Warning: Temperature below allowed working range Observe the lower working range limit
Bit 7 Warning: Temperature above allowed working range Observe the upper working range limit
Bit 8 Error: T  sensing element defective

Return the faulty unit to E+E for service

Bit 9 Warning: Moisture below allowed working range

Observe the lower working range limit

Bit 10 Warning: Moisture above allowed working range

Observe the upper working range limit

Bit 11 Error: Moisture sensing element defective

Return the faulty unit to E+E for service

Bit 12 Error: Hardware moisture sensing

Return the faulty unit to E+E for service

Содержание Modbus MOP301

Страница 1: ...BA_MOP301 v1 0 Modification rights reserved User Manual MOP301 Digital Moisture in Oil Immersion Probe up to 120 C 248 F ...

Страница 2: ... equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the in...

Страница 3: ... 1 Installation of the Probe Directly in the Process 8 4 2 Probe Installation with Ball Valve 9 4 3 Sampling Cell with Shut off Function 10 5 Setup and Adjustment 11 5 1 PCS10 Product Configuration Software 11 5 2 Modbus Setup 12 5 3 Modbus Register Map 13 5 4 Freely Configurable Custom Modbus Map 13 5 5 Device Status Indication 14 5 6 Modbus RTU Examples 15 6 Maintenance and Service 17 6 1 Cleani...

Страница 4: ...r to reach optimal performance of the device 1 2 Safety Instructions 1 2 1 General Safety Instructions Avoid any unnecessary mechanical stress and inappropriate use When replacing the filter cap make sure not to touch the sensing elements The device must be operated with the filter cap on at all times For sensor cleaning and filter cap replacement please see Cleaning Instructions at www epluse com...

Страница 5: ...understood this user manual and must follow the instructions contained within All process and electrical connections shall be thoroughly checked by authorized staff before putting the device into operation Do not install or start up a device supposed to be faulty Make sure that such devices are not accidentally used by marking them clearly as faulty A faulty device may only be investigated and pos...

Страница 6: ...il The MOP301 is typically deployed in monitoring of lubrication hydraulic and insulation oils as well as diesel fuel in equipment and machinery The oil to be monitored may be mineral synthetic or biodegradable Installation is simplified by various mounting options which are supported by E E accessories The MOP301 provides the measured data at its digital RS485 interface via Modbus RTU protocol Th...

Страница 7: ...10 pressure tight up to 20 bar 290 psi 3 3 Electrical Connection Important note The manufacturer cannot be held responsible for personal injuries or damage to property as a result of incorrect handling installation wiring power supply and maintenance of the device The relations of electrical potential between a bus system and an MOP301 probe are characterized by the following properties The bus co...

Страница 8: ...at precautions are taken to prevent the sensor from shooting out The internal ducting to the electronics is suitable for up to 20 bar 290 psi The respective feedthrough determines the actual working pressure 20 bar 290 psi As a rule non cross section constricting and cross section constricting cut in screw connections may be used cross section constricting screw connections are only permitted in t...

Страница 9: ... The temperature during probe installation may deviate by max 40 C 72 F from the regular temperature during normal operation Probe installation steps see Fig 6 Install the probe into the ball valve while the ball valve is closed Open the ball valve Slide the probe through the ball valve to the desired immersion depth Depending on the process pressure additional tools may be necessary for pushing t...

Страница 10: ...on works like the method described in chapter 4 1 Installation of the Probe Directly in the Process however without the need of stop valves in the upstream and downstream of the measurement point Technical data of the sampling cell1 Pipe thread Rp Whitworth thread according to EN 10226 old DIN 2999 Senor port thread G 1 2 ISO Operating pressure 40 bar at max 80 C 580 psi at max 176 F Temperature r...

Страница 11: ...hapter 5 1 below 5 1 PCS10 Product Configuration Software The PCS10 provides a convenient graphical user interface to the MOP301 To use the software for performing adjustments and changes in settings please proceed as follows 1 Download the PCS10 Product Configuration Software from www epluse com pcs10 and install it on the PC 2 Connect the MOP301 to the PC using the Modbus configuration adapter 3...

Страница 12: ...its can be set via 1 PCS10 Product Configuration Software and the Modbus configuration adapter HA011018 The PCS10 can be downloaded free of charge from www epluse com pcs10 2 Modbus protocol in the register 1 0x00 and 2 0x01 See Application Note Modbus AN0103 available at www epluse com mop301 The serial number as ASCII code is located in read only registers 1 8 0x00 0x07 16 bits per address The f...

Страница 13: ...and B 1663 30005 and 7 3705 decimal 46 10 00 64 00 04 08 E9 9A C4 CF DB 23 40 EB CA 19 If two parameters are to be uploaded it is recommended to write them with a single command Tab 5 MOP301 FLOAT32 measured data registers INT16 read register Parameter Unit Scale3 Register number1 Dec Register address2 Hex Read register function code 0x03 0x04 Water activity aw 1000 4068 0xFE3 Water content x ppm ...

Страница 14: ...t to NaN according to IEEE754 for data type FLOAT32 or to 0x8000 INT16 It is possible to read out all status and error information via Modbus register 602 0x259 Errors are displayed in bit coded form If an event is present the corresponding bit is set to 1 Measured values outside the measureing range are limited with the corresponding limit value Error Bits Description Recommended action Bit 0 Err...

Страница 15: ...E754 refer to the example at the end of this chapter Read register from custom modbus map address 0x0BB8 0x0BB9 unconfigured Request Hex Modbus address Function code Starting address Hi Starting address Lo Qty of registers Hi Qty of registers Lo CRC 46 03 0B B8 00 02 49 7D Response Hex Modbus address Function code Byte count Register 1 value Hi Register 1 value Lo Register 2 value Hi Register 2 va...

Страница 16: ...de Byte count Register 1 value Hi Register 1 value Lo Register 2 value Hi Register 2 value Lo Register 3 value Hi 46 03 0C 0B C9 41 DB BD Register 3 value Lo Register 4 value Hi Register 4 value Lo Register 5 value HI Register 5 value Lo Register 6 value Hi Register 6 value Lo CRC 7B 3E AD FF FF FF FF B7 9A Decoding of floating point values Floating point values are stored according IEEE754 The by...

Страница 17: ...he accuracy of a measurement device The device under test specimen is compared with the reference and the deviations are documented in a calibration certificate During the calibration the specimen is not changed or improved in any way Adjustment improves the measurement accuracy of a device The specimen is compared with the reference and brought in line with it An adjustment can be followed by a c...

Страница 18: ...PCS10 Modbus configuration adapter HA011018 Stainless steel filter cap for flow 1 m s HA010110 Stainless steel filter cap for flow 1 m s HA010111 Protection cap for M12 socket HA010781 Protection cap for M12 plug HA010782 T wcoupler M12 M12 HA030204 Ball valve G 1 2 ISO HA050101 Ball valve 1 2 NPT HA050104 Sampling cell with shut off function PN40 DN25 HA050109 ...

Страница 19: ...n rating IP66 NEMA 4X Pressure rating 20 bar 290 psi Probe material Stainless steel 1 4404 Cable jacket3 HFS 125XL black oil and fuel resistant Temperature working range Sensing element filter cap 40 125 C 40 257 F Probe 40 120 C 40 248 F Cable 40 120 C 40 248 F M12 connector 25 90 C 13 194 F Storage conditions 40 80 C 40 176 F 0 95 RH non condensing Electromagnetic compatibility EN 61326 1 EN 613...

Страница 20: ...epluse fr E E Elektronik Germany Obere Zeil 2 61440 Oberursel Tel 49 6171 69411 0 E mail info epluse de E E Elektronik India 801 Sakhi Vihar Road 400072 Mumbai Tel 91 990 440 5400 E mail info in epluse com E E Elektronik Italy Via Alghero 17 19 20128 Milano MI Tel 39 02 2707 86 36 E mail info epluse it E E Elektronik Korea Suite 2001 Heungdeok IT Valley Towerdong 13 Heungdeok 1 ro Giheung gu 16954...

Отзывы: