14
User Manual MOP301 Digital Moisture in Oil Immersion Probe
The register block for the configuration of the customisable Modbus map consists of the registers 6001
(0x1770) to 6010 (0x1779). For the blockwise query of the measured values behind Modbus registers 3001
(0xBB8) to 3020 (0xBCB), the firmware accesses this configuration area and thus gets the information
which measured value/status registers are to be output. A maximum of 10 user-defined registers can be
mapped. The table below shows an example:
Registers ...
... with these
assigned measurands ...
... map to registers ...
... mirrored from
source registers
Dec
Hex
Meas.
Unit
Type
Dec
Hex
Dec
Hex
Function code 0x10
Function code 0x03 / 0x04
6001
0x1770
Status
-
INT16
3001
0xBB8
602
0x259
6002
0x1771
T
°C
FLOAT32
3002
0xBB9
1003
0x3EA
°C
FLOAT32
3003
0xBBA
1004
0x3EB
6003
0x1772
T
°F
FLOAT32
3004
0xBBB
1005
0x3EC
°F
FLOAT32
3005
0xBBC
1006
0x3ED
6004
0x1773
T
°C
INT16
3006
0xBBD
4002
0xFA1
6005
0x1774
aw
-
FLOAT32
3007
0xBBE
1035
0x46E
-
FLOAT32
3008
0xBBF
1036
0x46F
6006
0x1775
x
ppm
FLOAT32
3009
0xBC0
1141
0x474
ppm
FLOAT32
3010
0xBC1
1142
0x475
6007
0x1776
Saturation
%
FLOAT32
3011
0xBC2
1137
0x470
%
FLOAT32
3012
0xBC3
1138
0x471
6008
0x1777
aw
-
INT16
3013
0xBC4
4068
0xFE3
6009
0x1778
x
ppm
INT16
3014
0xBC5
4071
0xFE6
6010
0x1779
Saturation
%
INT16
3015
0xBC6
4069
0xFE4
3016
0xBC7
65536
0xFFFF
3017
0xBC8
65536
0xFFFF
3018
0xBC9
65536
0xFFFF
3019
0xBCA
65536
0xFFFF
3020
0xBCB
65536
0xFFFF
Tab. 7 Custom Modbus map example
5.5 Device Status Indication
If a critical error occurs, all Modbus values are set to NaN (according to IEEE754 for data type FLOAT32)
or to 0x8000 (INT16). It is possible to read out all status and error information via Modbus register 602
(0x259). Errors are displayed in bit-coded form. If an event is present, the corresponding bit is set to 1.
Measured values outside the measureing range are limited with the corresponding limit value.
Error Bits Description
Recommended action
Bit 0 Error: Hardware T-sensing
Return the faulty unit to E+E for service
Bit 1 Error: Short circuit of T sensing element
1. Clean sensing head acc. to cleaning instructions
2. Return the faulty unit to E+E for service
Bit 2 Error: Open loop of T sensing element
Return the faulty unit to E+E for service
Bit 3 Error: Short circuit of moisture sensing element
1. Clean sensing head acc. to cleaning instructions
2. Return the faulty unit to E+E for service
Bit 4 Error: Open loop of moisture sensing element
or heavy pollution
1. Clean sensing head acc. to cleaning instructions
2. Return the faulty unit to E+E for service
Bit 5 Warning: Polluted moisture sensing element or
highly conductive oil
Clean sensing head or oil is incompatible
Bit 6 Warning: Temperature below allowed working range Observe the lower working range limit
Bit 7 Warning: Temperature above allowed working range Observe the upper working range limit
Bit 8 Error: T sensing element defective
Return the faulty unit to E+E for service
Bit 9 Warning: Moisture below allowed working range
Observe the lower working range limit
Bit 10 Warning: Moisture above allowed working range
Observe the upper working range limit
Bit 11 Error: Moisture sensing element defective
Return the faulty unit to E+E for service
Bit 12 Error: Hardware moisture sensing
Return the faulty unit to E+E for service