DYSIS Ultra 2.0 Digital Colposcope Instructions for Use
0230-53095, Rev 05
Revision Date 05-Mar-2021
Page
73
of
79
The EMR tab enables the user to allow DYSIS Ultra 2.0 Digital Colposcope
to interact with a site’s EMR system
in order to download patient demographic details or clinic appointments.
a.
DICOM Storage
To Configure DYSIS Ultra 2.0 Digital Colposcope System to Store Images to the PACS, the following settings
should be applied:
Local AE Title: DYSIS_COLPOSCOPE
Storage: Enabled
AE Title: DYSIS_PACS
Address: IP Address (the IP address of the PACS server)
Port: 11112
The Click “Check Connection” To make sure
everything works.
After this after a New Exam has been performed all Images should be sent to the PACS.
b.
Modality worklist
To configure DYSIS Ultra 2.0 Digital Colposcope system to use modality worklists, the following setting should
be applied:
Local AE Title: DYSIS_COLPOSCOPE
MWL: Enabled
AE Title: DYSIS_PACS
Address: IP Address (the IP address of the PACS server)
Port: 11112
To add an Item to the Modality Worklist for DYSIS Ultra 2.0 Digital Colposcope, an ORM^001 (Order Request
Message) should be sent to the DYSIS_PACS.