18.5.
Preparing the patient
Follow standard clinical practice to insert the speculum. Adjust the speculum to achieve the optimal field of
view. Gently clear any mucous using a dry swab or saline. Do not clean with acetic acid since it will start the
acetowhitening process. Ensure the speculum coupling adapter is positioned posteriorly.
DYSIS Ultra 2.0 Digital Colposcope can be connected to the speculum via the speculum coupling adapter at the
front of the optical head after insertion to stabilise the image during the examination. It is connected by
pushing the end of the speculum connecting rod into the opening on the green coupling adapter from the
Optical head. This does not lock into place but will stabilise the image whilst the examination takes place.
The Clinician monitor is a touch screen. From the DYSIS exam screen, the magnification can be increased
or decreased by using the
or by pinching or expanding the screen to zoom with your fingertips.
Examination Screen