198-1600-00 Rev F
3)
From the “Unit Start System Menu” rotate the current encoder until the display shows
“Strt” “P.FLt”, and then press the voltage encoder switch to view and change the power
failure alarm enable/disable setting.
4) The power failure alarm enable/disable setting is indicated by displaying
“Por FAIL FLt”
on the top display and the power failure alarm enable/disable setting in the bottom
display. The SET lamp is blinking on and off to indicate that a setting is displayed.
Rotate the current encoder to change the power failure alarm enable/disable setting.
Select “on” to enable power failure alarms. Select “OFF” to disable power failure
alarms. Once the desired configuration is selected press the voltage encoder switch to
accept the value and return to the “Unit Start System Menu”.
5)
From the “Unit Start System Menu” rotate the current encoder until the display shows
“Strt” “End”. Then press the voltage encoder switch to exit the “Unit Start System
Menu” and return to the “System Configuration Menu”.
6)
From the “System Configuration Menu” rotate the current encoder until the display
shows “SyS” “End”. Then press the voltage encoder switch to exit the “System
Configuration
Menu” and return to the “Unit Configuration Menu”.
7) If any settings were changed
in the “System Configuration Menu” you will be prompted
to store the changes. Select “yES” to accept the changes or “no” to cancel the
changes.
6.3.6. Remote Operate/Standby Control (Digital Input)
A digital input can be used to remotely control the operate/standby state. See
for details on wiring the remote operate control digital input.
The remote operate control digital input configurations are described in the table below. The “on
1” and “on 2” configurations are typically used for a momentary switch contact, such as a push
button switch, although they could also be used with a relay or other switch device controlled by a
PLC or other device.
Digital Input Operate Control Configuration
:
“d.in OPEr CtL” = “OFF”
The operate control digital input is disabled.
“d.in OPEr CtL” = “on”
The power supply operates when the operate control
digital input contacts are closed. Opening the
operate control contacts while the unit is operating
will put the unit into standby mode. If the contacts
are opened while a timed cycle (RTC or ATC) is
running the cycle is paused. If the contacts are
closed while a timed cycle is paused the cycle will
resume. Opening the contacts while the end of cycle
state is active (as indicated by “EOC” on the display)
will clear the end of cycle state. Note that the
operate state can be changed by the front panel
operate switch or the host port at any time (see
panel operate switch).
“d.in OPEr CtL” = “on 1”
The operate state is toggled each time the operate
input contacts are closed. After a timed cycle (RTC
or ATC) completes the end of cycle state is cleared