Dwyer Instruments UF35 Скачать руководство пользователя страница 1

DWYER INSTRUMENTS, INC.

Phone: 219/879-8000

www.dwyer-inst.com

P.O. Box 373 • Michigan City, IN  46361-0373, U.S.A.

Fax: 219/872-9057

e-mail: [email protected]

The Portable Series UF Sonicflow Flowmeters 

are non-inva-

sive “transit time” ultrasonic flowmeters designed to help ser-
vice/maintenance and commissioning engineers to take quick,
accurate flow readings.  Flow rates from 1.6 to 26 ft/sec (0.5 to 8
m/sec) of almost any clear or cloudy liquid can be measured with
±2% accuracy.  Easy to install flow transducers mount directly to
a variety of pipe materials such as stainless steel, cast iron, duc-
tile iron, PVC, copper, and concrete with diameters of 2˝ to 15.75˝.

FAST TRACK SETUP PROCEDURE

Switch on and press ENTER.  
Select Quick Start- Press ENTER.
Dimension Units?- Scroll to select units required, press ENTER.
Pipe OD- Enter data, press ENTER.
Pipe Wall Thickness- Enter data, press ENTER.
Pipe Lining Thickness- Enter data, press ENTER.  Enter Zero if
there is no lining on the application.
Select Wall Material- Select using scroll keys, press ENTER.
Select Lining Material- This will only be displayed if a lining thick-
ness has been entered.  Select using scroll keys, press ENTER.
Select Fluid Type- Select using scroll keys, press ENTER.

This instrument selects the mode of operation using the data
entered and will display the following:

Attach Sensor set

in XXXX mode

Approx. max. flow:

XXX m/s

Press ENTER to continue

Or SCROLL to change mode

Fluid Temp? Press Enter to input the application temperature in
the units required °C or °F then press ENTER.

Now retract the sensor blocks back into the guide rail by turning
the locking nuts clockwise. 
Apply grease to both sensor blocks as shown in (Figure 1),
attach to the pipe using the appropriate mounting hardware in
either Reflex or Diagonal Mode.  Ensure the Guide rail itself is
free of grease.

Figure 1
Inverted view
of grease
applied to sen-
sor block.

Model UF35

Specifications and Operating Instructions

Bulletin  F-UF35

SPECIFICATIONS

Range:

1.6 to 26 ft/sec (0.5 to 8 m/s).

Pipe Range:

2˝ to 15.75˝ (50 to 400 mm) nominal bore.

Accuracy:

± 2% of reading.

Repeatability:

± 0.5% or ± 0.1 LPM (whichever is greater).

Display: 

7-digit flow rate in GPM, LPM, LPS, m

3

/s, ft/sec and

m/s; 7-digit totalization in gallons, liters, or m

3

Display Resolution: 

0.1% of reading.

Response Time: 

< 2 seconds.

Operating Temperature: 

32 to 113°F (0 to 45°C).

Storage Temperature:

-4 to 131°F (-20 to 125°C).

Transducer Operating Temperature:

-4 to 257°F (-20 to

125°C).

Power: 

Four AA rechargeable batteries.

Guide Rail Length:

15.6˝ x 1.3˝ x 1.7˝ (395 x 33 x 42 mm).

Cable Length: 

6.5 feet (2 m).

Housing: 

ABS, IP55 rating.

Weight:

17.5 oz (500 g) including transducers.

Connect the red and blue sensor cables to the electronics and
the guide rail assembly.  The RED cable must be positioned
upstream to give a positive flow reading.

Figure 2- Sensor
Assembly

For Reflex Mode attach the guide rail (Figure 3) to the pipe as
shown below.  Turn the locking nut on the fixed transducer anti-
clockwise, screwing it down on to the pipe surface.  No not over-
tighten, causing the guide rail to lift off the pipe.
Set the separation distance (Figure 3) by sliding the floating trans-
ducer along the scale until the front edge of the block is at the
recommended distance displayed by the electronics.  Now turn
the locking nuts on both the floating and the fixed transducers
anti-clockwise, until they make finger tight contact with the pipe
surface.  To mount the transducers in DIagonal Mode, refer to
(Figure 4).

Now press ENTER to read flow.  Pressing the appropriate key on
the keypad can change flow units.  An additional key press will
change the time scale of the reading- hr/min/sec.

Figure 3: Reflex Mode Operation

UF35 bulletin  6/13/03  10:50 AM  Page 1

Содержание UF35

Страница 1: ... in either Reflex or Diagonal Mode Ensure the Guide rail itself is free of grease Figure 1 Inverted view of grease applied to sen sor block Model UF35 Specifications and Operating Instructions Bulletin F UF35 SPECIFICATIONS Range 1 6 to 26 ft sec 0 5 to 8 m s Pipe Range 2 to 15 75 50 to 400 mm nominal bore Accuracy 2 of reading Repeatability 0 5 or 0 1 LPM whichever is greater Display 7 digit flow...

Страница 2: ... for m s press 4 again for f s Press 7 for l s press 7 again for l m Press 8 for g min press 8 again for USG min Press 9 for m3 min press 9 again for m3 sec There are some facilities that require the cursor to be moved from left to right This can be done using keys 5 left and 6 right The pulse output can only be activated in the flow mode Transducers The UF35 is supplied with one matched pair of t...

Страница 3: ...ment to READ FLOW Main Menu View Edit Site Data The VIEW EDIT SITE DATA mode can be accessed from the main menu It allows the user to enter application details for up to 20 different sites This facility is useful if a number of sites are being monitored on a regular basis Application data can be pro grammed into each site before getting to the site When scrolling up down the menu press ENTER to se...

Страница 4: ...s use enhanced mode when signals are below 800 Below 800 the system may gener ate noise and therefore the accuracy of the measurement cannot be guaranteed Sensor Parameters This facility is passwords protected It stores sensor information used by Dwyer and is not available for the user Factory Settings The facility is used by Dwyer in the process of instrument calibra tion Pressing ENTER takes the...

Страница 5: ... µs below the Up µs Dn µs value Phase up dn µs Only valid if Calculated µs and Up µs Dn µs are correct If the reading is zero then there is no signal which could mean the pipe is empty or the liquid is contaminated with particles or air Phase offset This value will be between 0 and 15 The exact value is not important and will vary between applications It should however be stable when the flow cond...

Страница 6: ...nsor set CANNOT READ FLOW BECAUSE Sensor mode is invalid for this pipe size Temperature range is 20 C to 125 C The temperature range of the transducers is 20 to 125 C Enter a lining thickness first This message appears when in VIEW EDIT SITE DATA the user has tried to enter a pipe lining material before entering a thick ness APPLICATION INFORMATION The UF35 is a Transit Time ultrasonic flow meter ...

Страница 7: ...ut when the transducers are sighted this close to any obstruction errors can be considerable It is not possible to predict the amount of error as this depends entirely upon the type of obstruction and the configuration of the pipe work and flow profile The message therefore is clear Do not expect to obtain accurate results if the transducers are positioned closer than allowed to any obstruction th...

Страница 8: ...he UF35 has been calibrated to operate on turbulent flows with a Reynolds Number of approximately 100 000 The calibration of the unit will not be valid if the Reynolds No is below 4000 If the UF35 is to be used on laminar flow applications it will be necessary to calculate the Reynolds No for each application To calculate the Reynolds No it is necessary to know the Kinematic viscosity in Centistok...

Страница 9: ...UE sensor cable downstream The electronics will display a positive flow reading with cables in this orientation If the unit displays a nega tive reading the cables have been connected to the wrong sen sors MAINTENANCE Following installation the UF35 requires no routine maintenance Aside from methods described in this manual units are not field serviceable and should be returned to the factory if r...

Отзывы: