Dwyer Instruments DPGA Series Скачать руководство пользователя страница 1

DWYER INSTRUMENTS, INC.

Phone: 219/879-8000

www.dwyer-inst.com

P.O. Box 373 • Michigan City, IN  46361-0373, U.S.A.

Fax: 219/872-9057

e-mail: [email protected]

The Series DPGA & DPGW Digital Pressure Gages

have a p/- 1.0% full scale accuracy. The 4 digit
digital display will reduce the potential for errors in readings
by eliminating parallax error commonly produced with
analog gages.

The DPGA & DPGW are packaged in an ABS plastic case.
The unit is battery powered and has an auto-shut off to
conserve battery life. A two button key pad allows easy
access to features without the need to work through
complex menus. These features include, auto zero and
conversion of the pressure units. Readings may be
converted to various engineering units. See reverse page
for available units.

OPERATING INSTRUCTIONS

For complete instructions on the two button operation see
the flow chart provided on back.

On/Off:

Press ON/OFF to toggle on and off. Device will

automatically shut off after 20 minutes if not used.

Zero:

Zeros the display - Push & hold for 2 seconds to zero

the display. If after pressing zero the display does not re-
zero, the unit is no longer meeting specified accuracy and
should be sent back to the factory for calibration. Do not
use the zero button when pressure is applied.

Units:

To change units, press both On/Off and Zero

buttons simultaneously.

Overpressure Indicator:

The LCD will flash “OFL”, if the

pressure applied is over 105% F.S.

INSTALLATION

When installing gage always use 1˝ hex at the base of the
housing to tighten the gage to a mating fitting. Do not apply
wrench to housing.

MAINTENANCE

Battery Removal:

Remove the screw on backplate. Lift

backplate off by hand. To reassemble replace backplate
and screw.

Replace batteries per polarity indicators.

A “LOW BAT” descriptor indication will appear on the
display when batteries need to be replaced.

Series DPGA & DPGW Digital Pressure Gages

Specifications - Installation and Operating Instructions

Bulletin  A-34C

Ø2-43/64 (67.87)

3-11/16

(93.66)

1-1/2 (38.10)

1/4 NPT

SPECIFICATIONS
Service:

DPGA: Air and compatible gases; DPGW:

Compatible gases and liquids.

Wetted Materials: 

DPGA: 316L SS, silicone sensor;

DPGW: 316L SS.

Housing Materials:

ABS plastic.

Accuracy:

±1.0% F.S. (Includes linearity, hysteresis,

repeatability). 

Pressure Limits: 

2X pressure range. Vacuum range max

pressure is 30 psig.

Temperature Limits: 

30 to 120°F (-1 to 49°C).

Thermal Effect: 

0.05% FS/°F.

Size:

2.62˝ O.D. x 1.52˝ deep.

Process Connections: 

1/4˝ male NPT.

Display:

4-digit LCD (.425˝ H x .234˝ W digits).

Power Requirements: 

9 volt alkaline battery. Battery

included but not connected.

Auto Shut-off: 

20 minute auto shut-off.

Weight: 

5.6 oz (160 g).

Agency Approvals:

CE.