Dwyer Instruments 477B Series Скачать руководство пользователя страница 1

Series 477B Handheld Digital Manometer

Bulletin D-26B

Specifications - Installation & Operating Instructions

SPECIFICATIONS
Service: 

Air and compatible gases

Wetted Materials: 

Consult factory

Accuracy:

±0 1% F S  from 60 to 78°F (15 6 to 256°C)  ±1% F S  from 32 to 60

and 78 to 104°F (0 to 15 6 and 25 6 to 40°C)

Pressure Hysteresis: 

±0 1% F S

Pressure Limits: 

See chart

Temperature Limits: 

Operating  0 to 140°F (-17 8 to 60°C)  Storage  -4 to 176°F

(-20 to 80°C)

Display: 

4-digit LCD ( 425˝ H x  234˝ W digits)

Resolution:

See chart

Process Connections: 

Two barbed connections for use with 1/8˝ (3 18 mm) or

3/16˝ (4 76 mm)  D  tubing for 477B-1  477B-2  477B-3  477B-4  and 477B-5 only
Two compression fittings for use with 1/8˝ (3 18 mm)  D  x 1/4˝ (6 35 mm) O D
tubing for 477B-6 and 477B-7 only

Power Requirements: 

9 V alkaline battery  installed non-functional  user

replaceable

Weight: 

10 2 oz (289 g)

Agency Approvals: 

CE

BACKLIGHT

The Model 477B includes a display backlight to allow use in the dark or in poor
lighting conditions  The manometer must be switched off before this feature can be
activated Next press and hold the ON/OFF button After about one second the
backlight will come on and remain lighted for approximately 2 minutes after which it
will turn itself off to conserve battery life

ZEROING PRESSURE READING

Potential inaccuracy due to temperature effects can be minimized by re-zeroing
immediately before use  To zero the display  vent both ports to the atmosphere so
no pressure is applied to either port  Press the ZERO/STORE button and ---- will be
momentarily displayed as zeroing occurs  Zeroing is not possible when the memory
mode is in use   t must be done before selecting that function

f  the  unit  is  accidentally  zeroed  with  pressure  applied  to  one  of  the  ports the

pressure reading might display incorrectly  To correct vent the pressure ports to
atmosphere and press the ZERO/STORE button to zero the unit

The Series 477B Handheld Digital Manometer 

is a versatile  handheld  battery

operated manometer available in several basic ranges from 0-20 in w c  up to 100
psi All models measure either positive negative or differential pressures with
±0 1% of full scale accuracy  Users can select from up to seven common English
and metric pressure units  so conversions are not necessary  A memory function
allows storage of up to 40 readings for later recall and a backlight provides auxiliary
lighting in locations where it is too dark to read the measurements on the unit

BATTERY INSTALLATION

The unit is shipped with a separate 9 volt alkaline battery which must be installed
before operation  Remove the two screws holding the bottom endcap in place and
remove the endcap  Connect the battery to the enclosed battery clip  making sure
to observe correct polarity Be careful not to trap wires between the battery  case
or foam pads which retain the battery  This could make it difficult to install the battery
or remove it later for replacement  Be sure that the rubber gasket is properly seated
in the gasket channel of the endcap and replace endcap  Note that the endcap will
only fit one way because the holes are slightly off-center  Place the Z shaped wrist
strap clip in one of the screw recesses and replace the screws  Do not over tighten
the screws  Attach the wrist strap to the clip

When battery replacement becomes necessary  use only a 9 volt alkaline type such
as a Duracell

®

MN 1604  Eveready

®

522  or equivalent  Zinc-carbon types  often

labeled heavy-duty are not recommended because of the increased potential for
leakage Alkaline batteries are also a better value because they last up to three
times longer in this device

Low Battery Indicator

A weak battery can cause improper operation or inaccurate measurements A low
battery  indicator  is  provided  on  the  display  to  show  when  the  battery  needs
replacement  

ON-OFF OPERATION

The on-off control is a toggle function  Press and release the ON/OFF button once
to turn the unit on and press it again to turn it off f the manometer is left on with no
activity for approximately 20 minutes the unit will turn itself off to conserve the
battery

41/64

[16.27]

1-1/16

[26.99]

BARBED FOR 1/8
OR 3/16 ID TUBING

5-57/64

[149.68]

2-13/16

[71.44]

29/32

[23.02]

Although the unit might appear to function and indicate properly
the accuracy of readings cannot be guaranteed when the LOW

BAT indicator is illuminated  Replace the battery with a new one  Do not leave an
exhausted battery in the unit due to potential leakage

NOTICE

Model 
477B-1
477B-2
477B-3
477B-4
477B-5
477B-6
477B-7

English 
Range

0 to 20 in w c
0 to 40 in w c
0 to 200 in w c
0 to 10 psi
0 to 30 psi
0 to 50 psi
0 to 100 psi

Metric
Range

0 to 4 982 kPa
0 to 9 96 kPa
0 to 49 82 kPa
0 to 68 95 kPa
0 to 206 9 kPa
0 to 344 8 kPa
0 to 689 5 kPa

Available Pressure Units

psi  in w c in Hg  mm w c mm Hg  Pa kPa bar  mbar
psi  in w c in Hg  mm w c mm Hg  Pa kPa bar  mbar
psi  in w c in Hg  mm w c mm Hg  Pa kPa bar  mbar
psi  in w c in Hg  mm w c mm Hg  Pa kPa bar  mbar
psi  in w c in Hg  mm Hg  kPa bar  mbar
psi  in w c in Hg  mm Hg  kPa bar  mbar
psi  in w c in Hg  mm Hg  kPa bar  mbar

Maximum
Pressure

3 psi (0 21 bar)
3 psi (0 21 bar)
15 psi (1 03 bar)
30 psi (2 07 bar)
60 psi (4 13 bar)
100 psi (6 89 bar)
200 psi (13 78 bar)

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

 TestEquipmentDepot.com 

Отзывы: