Selecting Thermocouple Input Type

Step 1: While the meter is off, press and hold the        and          buttons until the
current thermocouple type shows on the left side of the display.   

Step 2: Press the           button to toggle between K, J and T thermocouple types.

Step 3: Press the          button to save the selection. An “S” will appear on the
display to show that the setting has been saved.

Max and Min Data Recording

Press the            button momentarily and the “REC” icon will appear on the display.
While the “REC” icon appears on the display, the minimum and maximum readings
are recorded for the duration of time the meter is in record mode.

Press the          button in order to display the maximum temperature recorded.
Press again in order to display the minimum temperature recorded.

To return to normal operation, press and hold the           button for 3 seconds.

Note:

When the            button is pressed, all other buttons are disabled except the

power and backlight buttons.

Automatic Power Off (Sleep Mode)

The  instrument  will  shut  off  automatically  after  20  minutes  of  use  to  conserve
battery life if no button is pressed.

To disable the sleep mode, press        and the             buttons simultaneously before
power on. An “n” icon will appear in the center of the display once disabled.

OPERATING INSTRUCTIONS
Taking Measurements

Step 1: Before the meter is turned on, insert any J, K, or T type thermocouple(s)
that has a mini plug connection into the top of the meter.

Step 2: Press the        button to turn the meter on.

Step 3: After three seconds, the primary larger display will show the value for Probe
1 and the secondary display will show the value for Probe 2.

Step  4:  If  viewing  in  a  dark  area,  press  the          button  momentarily  to  activate
backlight.

Changing Engineering Units

Press the            button to toggle between degrees Celsius and degrees Fahrenheit.

Selection of Display Measurements

Press the              button to switch the main display between Probe 1 measurement,
Probe 2 measurement, and the Differential Temperature (Probe 1 – Probe 2). The
secondary display will show the opposite probe temperature from the main display.
While the main display shows the Differential Temperature, the secondary display
will toggle between the values of Probe 1 and Probe 2.

Data Hold

Press  and  release  the                button  to  freeze  the  primary  temperature  display.
“HOLD” will appear at the top of the screen.

In order to return to normal operation, press the            button again and the “HOLD”
icon disappears from the display.

The Model 472A-1 Dual Input Thermocouple Thermometer

precisely measures

up  to  two  temperature  measurements  simultaneously.  The  large  LCD  display
shows  both  temperature  inputs  or  one  temperature  input  and  the  differential
temperature. Any J, K, or T type thermocouple with a mini-jack connector can be
used  as  an  input.  For  viewing  in  poorly  lit  environments,  the  built-in  backlight
brightens  the  display. A  hold  button  allows  the  user  to  freeze  temperature  data
displayed and minimum and maximum readings can be recorded over a set time
period.  Model  472A-1  includes  a  hard  carrying  case,  battery,  and  one  K  type
thermocouple.

Model 472A-1 Dual Input Thermocouple Thermometer

Specifications - Installation and Operating Instructions

Bulletin TE-472A

DWYER INSTRUMENTS, INC. 

Phone: 219/879-8000

www.dwyer-inst.com

P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A. 

Fax: 219/872-9057

e-mail: [email protected]

SPECIFICATIONS
Inputs: 

Type J, K, T thermocouples.

Power Requirements: 

9 V, alkaline battery, installed non-functional, user

replaceable.

Measurement Ranges: 

J-type: -328° to 1400°F (-200° to 760°C);
K-type: -328° to 2498°F (-200° to 1370°C);
T-type: -328° to 734°F (-200° to 390°C);

Accuracy: 

±0.1% r 1.4°F (0.7°C).

Operating Temperature: 

32° to 122°F (0° to 50°C).

Operating Humidity (Non-Condensing): 

0 to 85% relative humidity.

Display:

Triple LCD display.

Resolution:

0.1°C up to 500°C.

Weight:

1.47 lb.

Agency Approval: 

CE.

REC

REL

HOLD

°C/°F

CHANNEL

REL

HOLD

=Disable SLEEP MODE

REC

°C/°F

SELECT TYPE

REC

SAVE SELECTED TYPE.

RS

23

2

D

C

9V

OPEN

19/64

[29.08]

219/32

[65.79]

71/16

179.32

15/16

[33.52]

213/16

[71.56]

REL

HOLD

REL

HOLD

°C/°F

CHANNEL

REC

REC

REC

REC

REL

HOLD

REC

°C/°F

REC

Отзывы: