2.5  Multiple RAD7s 

Before using the Natural Rock Sample to check a RAD7’s radon sensitivity, the Natural Rock Sample 
must be le

ft

 sealed for a month, in order for the radon inside the container to reach equilibrium. 

Th

at 

means a RAD7 can have its sensitivity checked each month on a regular basis. However, there is 
nothing to stop us hooking up multiple RAD7s in series in the loop connected to the Natural Rock 
Sample. 

Th

is increases the volume of the loop and hence decreases the radon concentration in the loop, 

but by a predictable and reproducible amount. 

Th

is method is recommended for checking up to six 

RAD7s on a monthly basis. 

For the sake of argument, using rounded 

gures, let us assume that the total activity of radon inside the 

Natural Rock Sample container, when in equilibrium, is equal to 2 Bq. Let us also assume that the 
Natural Rock Sample and tubing has an air volume of 500ml while a RAD7 is 1L. 

Th

en when the RAD7 

is connected up to the Natural Rock Sample and the radon distributed around the loop, the radon 
concentration will be 2 / 1.5 = 1.333 Bq/L = 1333 Bq/m

3

. If every additional RAD7 adds 1L to the loop, 

then four RAD7s would total 4.5L, including the Natural Rock Sample and tubing. 

Th

e radon 

concentration in the loop would then become 2 / 4.5 = 0.444 Bq/L = 444 Bq/m

3

. While this is less than 

1333 Bq/m

3

, it is still enough to make a good reading in two hours and a high precision reading in 12 

hours. 

 

Figure 5  Multiple RAD7 Sensitivity Check Con

guration 

Each month, all four RAD7s can be connected in series with one another and thoroughly purged and 
dried out, before being connected to the rock sample to obtain a sensitivity check. Each month the 
setup should be identical, with the same number of RAD7s and the same pieces of tubing to connect up 
the entire loop. During the measurement all the RAD7 pumps, except one, should be set to OFF 
(

Setup,

 

Pump,

 

Off,

 

[ENTER]

.) 

If any of the RAD7s had developed a leak, the radon concentration in the loop would fall rapidly. 

Th

us 

in addition to keeping a record of the average readings during the check, for each month, a record of 

Sec$on 2

 

 

RAD7 Radon Sensitivity Check 

 

 

 

 

 

 

      

8

Содержание NATURAL ROCK SAMPLE

Страница 1: ...NATURAL ROCK SAMPLE Performance Monitoring Accessory for the RAD7 User Manual...

Страница 2: ...2 3 1 Drying and Resealing the Natural Rock Sample 5 2 4 Radon Data Assessment 6 2 5 Multiple RAD7s 8 2 6 Cumulative Spectrum 9 3 RAD7 THORON SENSITIVITY CHECK 10 3 1 Thoron Measurement 10 3 1 1 Radon...

Страница 3: ...the closed loop radon calibration circuit or upstream of the desiccant in the open loop thoron calibration con guration If the desiccant in the Small Drying Tube gets used up it should be replaced or...

Страница 4: ...be compared with the known concentration and thus the sensitivity of the RAD7 may be e ectively checked A correction may be applied to the Natural Rock Sample radon reading to account for the ambient...

Страница 5: ...Setup Protocol Recycle 06 ENTER Put the RAD7 s infrared printer in place and switch o the RAD7 Switch on the RAD7 printer Switch on the RAD7 and let it print out the header including a review of the s...

Страница 6: ...re one month from that date The next measurement should be made within 50 days of the date on the label to avoid the risk of moisture creeping in and changing the emanation e ciency and hence the rado...

Страница 7: ...ock Sample system with standard RAD7 of 1 275 L If Rn is expressed in pCi then the radon concentration will be Rn 1 275 pCi L If Rn is expressed in Bq then the radon concentration will be Rn 1 275 Bq...

Страница 8: ...al Rock Sample and the radon distributed around the loop the radon concentration will be 2 1 5 1 333 Bq L 1333 Bq m3 If every additional RAD7 adds 1L to the loop then four RAD7s would total 4 5L inclu...

Страница 9: ...d with all pumps except one set to OFF 7 Depending on the precision required after between 3 and 24 hours the tests may be stopped and the RAD7s disconnected 8 If required rough thoron checks may then...

Страница 10: ...measurement The setup assumed in the RAD7 data processing and stated as standard in the manual consists of a Small Drying Tube which may be used as a wand for sni ng along with a standard input tubing...

Страница 11: ...Sample it will make no di erence to the average radon reading but will make a big di erence to the thoron readings Even just switching the connections to the Natural Rock Sample can a ect the thoron r...

Страница 12: ...Figure 7 Thoron Sensitivity Check con guration Sec on 3 RAD7 Thoron Sensitivity Check 12...

Страница 13: ...n check the rate of injection of thoron into the sample ow will always be the same Let this be Th The units of Th are pCi per minute or Bq per minute If the RAD7 ow rate measured at the RAD7 outlet is...

Страница 14: ...978 667 9556 Fax 1 978 667 9557 Web www durridge com Email service durridge com Copyright 2020 DURRIDGE Company Inc All rights reserved DURRIDGE the DURRIDGE logo and the Works with the DURRIDGE logo...

Отзывы: