DURRIDGE NATURAL ROCK SAMPLE Скачать руководство пользователя страница 6

Figure 3  Natural Rock Sample drying loop con

guration 

Set the RAD7 pump to ‘ON’ (

Setup,

 

Pump,

 

On

 

[ENTER]

), go to third status window (

Setup, 

Enter,

 

Enter,

 

Rt. Arrow,

 

Rt. Arrow.

) and monitor the RH reading. Keep the air circulating 

through the loop until the relative humidity (RH) drops to 4% or below. 

Th

e drying process may take 

many hours. 

Once the RH stays at 4% or below, turn o

 the pump (

Setup,

 

Pump,

 

Off

 

[ENTER]

), close the ball 

valves, disconnect the tubing, and insert the end caps on the Natural Rock Sample tube. Mark the date 
on the label and store the Natural Rock Sample with its tubing and Small Drying Tube on a shelf for 
between 30 and 50 days before repeating the process. 

If the ball valves are opened for any reason on a later date, remove the old label and replace it with a 
new label showing the last date on which the valves were closed, and le

ft

 closed. 

Th

e rock sample will 

be ready for another radon check not before one month from that date. 

Th

e next measurement should 

be made within 50 days of the date on the label, to avoid the risk of moisture creeping in and changing 
the emanation e

ciency, and hence the radon and thoron output, of the Natural Rock Sample. 

2.4  Radon Data Assessment 

If the Natural Rock Sample was never calibrated, keep the average of the last four readings (3, 4, 5, and 
6) in a record for future reference. (Figure 4 on the next page shows a typical data set in DURRIDGE’s 
CAPTURE so

ft

ware, with the last four readings selected.) Every standard RAD7 should produce nearly 

the same result with that Natural Rock Sample if its sensitivity is as calibrated and if the rock sample is 
dry and has been sealed for between 30 and 50 days. For greater reproducibility, 2/3 of the ambient air 
radon concentration can be subtracted from the averaged readings to correct for the residual radon in 
the air in the system a

ft

er purging. However, if the ambient radon concentration in the air is less than 

100 Bq/m

3

 (2.7 pCi/L) the error introduced by ignoring it will be less than 1%. 

Sec$on 2

 

 

RAD7 Radon Sensitivity Check 

 

 

 

 

 

 

      

6

Содержание NATURAL ROCK SAMPLE

Страница 1: ...NATURAL ROCK SAMPLE Performance Monitoring Accessory for the RAD7 User Manual...

Страница 2: ...2 3 1 Drying and Resealing the Natural Rock Sample 5 2 4 Radon Data Assessment 6 2 5 Multiple RAD7s 8 2 6 Cumulative Spectrum 9 3 RAD7 THORON SENSITIVITY CHECK 10 3 1 Thoron Measurement 10 3 1 1 Radon...

Страница 3: ...the closed loop radon calibration circuit or upstream of the desiccant in the open loop thoron calibration con guration If the desiccant in the Small Drying Tube gets used up it should be replaced or...

Страница 4: ...be compared with the known concentration and thus the sensitivity of the RAD7 may be e ectively checked A correction may be applied to the Natural Rock Sample radon reading to account for the ambient...

Страница 5: ...Setup Protocol Recycle 06 ENTER Put the RAD7 s infrared printer in place and switch o the RAD7 Switch on the RAD7 printer Switch on the RAD7 and let it print out the header including a review of the s...

Страница 6: ...re one month from that date The next measurement should be made within 50 days of the date on the label to avoid the risk of moisture creeping in and changing the emanation e ciency and hence the rado...

Страница 7: ...ock Sample system with standard RAD7 of 1 275 L If Rn is expressed in pCi then the radon concentration will be Rn 1 275 pCi L If Rn is expressed in Bq then the radon concentration will be Rn 1 275 Bq...

Страница 8: ...al Rock Sample and the radon distributed around the loop the radon concentration will be 2 1 5 1 333 Bq L 1333 Bq m3 If every additional RAD7 adds 1L to the loop then four RAD7s would total 4 5L inclu...

Страница 9: ...d with all pumps except one set to OFF 7 Depending on the precision required after between 3 and 24 hours the tests may be stopped and the RAD7s disconnected 8 If required rough thoron checks may then...

Страница 10: ...measurement The setup assumed in the RAD7 data processing and stated as standard in the manual consists of a Small Drying Tube which may be used as a wand for sni ng along with a standard input tubing...

Страница 11: ...Sample it will make no di erence to the average radon reading but will make a big di erence to the thoron readings Even just switching the connections to the Natural Rock Sample can a ect the thoron r...

Страница 12: ...Figure 7 Thoron Sensitivity Check con guration Sec on 3 RAD7 Thoron Sensitivity Check 12...

Страница 13: ...n check the rate of injection of thoron into the sample ow will always be the same Let this be Th The units of Th are pCi per minute or Bq per minute If the RAD7 ow rate measured at the RAD7 outlet is...

Страница 14: ...978 667 9556 Fax 1 978 667 9557 Web www durridge com Email service durridge com Copyright 2020 DURRIDGE Company Inc All rights reserved DURRIDGE the DURRIDGE logo and the Works with the DURRIDGE logo...

Отзывы: