56
2
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
3
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
G
1
5. CHANGING CHEMICALS
Chemicals should be changed every 3
weeks.
Advice:
The 3 weekly chemical change
specifically refers to:
a) The use of
„DURR Automat XR
chemicals“.
b) A daily
max.
film throughput of
10
intra-oral
and
3 panoral
films.
In the BRD, as well as some other
countries, used x-ray chemicals are
regarded as special waste and
must be collected and disposed of
separately. We recommend the use
of a DURR Collection & Storage
system, see left fold-in page (3) and
(4).
5.1 Cleaning the roller sets
Advice
: Any hard particle deposits can
safely and easily be removed with
the special DURR Spray Cleaner
WR 200 (P/N CCB800C55..). For
Cleaning the roller racks periodically
(3-6 mos) please use DURR
Cleaning Bath set (P/N 1700-007-
00).
Left and right fold-in pages:
• Mains switch (G) off, see fig. 1
• Pull out mains plug.
• Lift off lid (7) and put to one side.
• Open all latches (25) and catches (24),
see right fold-in page.
• Remove guide plate (8) from dryer set.
• Tilt dryer set (23) backwards and then lift out.
• Remove water roller set (26) and film
entrance set (28).