9. Hose connections
9.1 Hose connection to XR24, XR 24 II,
XR 24 NOVA or to XR 24 PRO
with serial number up to T005999
The regeneration connection is
effected using the nozzles to the tank
at 16E and 16F, see figs. 6 and 7.
• The connecting hose from the Regeneration
Unit for developer (black) (13) and fixer
(red) (14) must be slipped through the
prepared holes on the rear of the machine
and fed to the grey connecting sleeves
(16E) and (16F) on the baths.
• Uncsrew the locking nuts (18) on the
connectors (16E) and (16F).
• Remove the black plugs (20). Slide the
locking nuts (18) and the grey hose
locking ring (19) onto the connecting hose
(13), and (14). Slide the connecting hose
for developer (black) (13) and fixer (red)
(14) all the way onto the appropriate
connector sleeve, (16E) and (16F), of the
bath and secure in place with the locking
nut (18).
E
Developer bath
F
Fixer bath
9.2 Hose connection to XR 24 PRO
with serial number from T006000
The tank design has been optimised
and the regeneration connection is
effected at the inlet pipe, see fig. 7a.
Use the assembly set 1700-284-00 “inlet
pipe for regeneration unit” with enclosed
assembly instructions.