Usage
Only use a cleaning agent that is compatible
with the materials:
Dürr Dental recommends the image plate clean-
ing wipe (see "Cleaning and disinfection"). Only
this product has been subjected to material
compatibility testing by Dürr Dental.
NOTICE
Heat or humidity will damage the im-
age plate.
i
Do not steam sterilise the image plate.
i
Do not immersion-disinfect the image
plate.
i
Only use cleaning agents that are
compatible with the materials.
i
Soiling on both sides of the image plate
should be cleaned off with a soft, lint-free
wipe prior to every use.
i
Remove resistant or dried on dirt with the im-
age plate cleaning wipe. When doing this, ob-
serve the instructions for use for the cleaning
wipe.
i
Allow the image plate to completely dry before
using it.
12.4 Protection cover
Clean the surface of the protective cover if it is
obviously dirty.
i
Clean the protective cover with a soft, lint-free
cloth that has been moistened with cold tap
water.
i
Only fit the protective cover to a unit that has
been cleaned and disinfected.
12.5 Storage box with image plate
storage tray
Clean and disinfect the surface of the storage
box and the internal image plate storage tray in
the event of contamination or visible soiling.
Dürr Dental recommends the following disinfect-
ant for the storage box:
FD 366 sensitive
Dürr Dental recommends the following disinfect-
ants for the image plate storage tray:
FD 350 and FD 366 sensitive
i
Clean the surface of the storage box and the
image plate storage tray with a soft, lint-free
cloth that has been dampened with cold tap
water.
i
Disinfect the storage box using a disinfection
wipe. Alternatively, use disinfectant on a soft,
lint-free cloth.
i
Clean the cover, fixing mechanism and inside
parts with a soft, lint-free cloth that has been
dampened with cold tap water.
i
Disinfect the cover, fixing mechanism and in-
side parts with a disinfection wipe.
Alternatively, use disinfectant on a soft, lint-
free cloth. Comply with the operating instruc-
tions for the disinfectant when doing this.
The cover and fixing mechanism can also
be disinfected in an immersion disinfection
system.
i
Remount the fixing mechanism.
i
Remount the cover.
i
Touch
OK
on the touch screen.
The fixing mechanism moves into the start-
ing position.
12.2 Light protection cover
The surface of the unit must be cleaned and
disinfected if it is contaminated or visibly soiled.
i
Disinfect the light protection cover using a dis-
infectant before and after placement. Dürr
Dental recommends FD 333 forte wipes (viru-
cidal), FD 350 (limited virucidal activity) and FD
322 premium wipes (limited virucidal activity).
i
Allow the light protection cover to completely
dry before using it.
12.3 Image plate
Cleaning and disinfection wipes are unsuitable
for cleaning image plates and may cause dam-
age to them.