9000-615-32/02 1808V002
37
Usage
Disinfect using Vector/RinsEndo Disinfection:
i
Pour 2 sealing caps (approx 40 ml) of
Vector/
RinsEndo disinfection
into the fluid container
2x
i
Touch the RINSE (rinsing/disinfecting) button
for at least 2 seconds
RINSE
i
LED flashes – the
Vector/RinsEndo Disinfec-
tion
is rinsed into the system and the process
ends automatically The
Vector/RinsEndo
Disinfection
remains in the system until the
next treatment
i
Empty any remaining
Vector/RinsEndo Disin-
fection
from the fluid container
i
Thoroughly rinse the fluid container with water
and dry
14.2 Initial start-up after a break
in treatment for more than
24 hours
The initial start-up process will depend upon
whether the hose system was reprocessed be-
fore the break in treatment Proceed as follows
depending on the situation:
1. Reprocessing performed before the treat-
ment break:
i
Rinse the system with water
Thorough rinsing with water will flush out
any disinfectant remaining in the hoses,
and any irritation to the patient caused by
the taste of residual disinfectant will be
avoided
i
Pull off the hose connection from the hand-
piece
i
Lay the handpiece hose in the sink
i
Fill the fluid container with water
i
Touch the RINSE (rinsing/disinfecting) button
for at least 2 seconds
LED flashes – the rinsing process takes
around 30 seconds and ends automatically
14 Treatment breaks for more
than 24 hours
If no treatment is carried out for a period of
24 hours or more, reprocessing of the hose sys-
tem must be performed
14.1 Cleaning and disinfecting the
hose system
The hose system is disinfected using a ready-
to-use, aldehyde-free solution of
Vector/Rin-
sEndo Disinfection
i
Pull off the hose connection from the hand-
piece
i
Lay the handpiece hose in the sink
Rinsing with water:
i
Fill the fluid container to approx 1/3 with wa-
ter
i
Touch the RINSE (rinsing/disinfecting) button
for at least 2 seconds
RINSE
i
LED flashes – the rinsing process takes
around 30 seconds and ends automatically
i
Rinsing with water will flush out any remaining
blockage
i
After the end of the rinsing process drain any
remaining fluid in the system
EN