![Durr Dental Vector Scaler Скачать руководство пользователя страница 38](http://html.mh-extra.com/html/durr-dental/vector-scaler/vector-scaler_installation-and-operating-instructions-manual_2542917038.webp)
36
9000-615-32/02 1808V002
Usage
Requirements placed on the steam steriliser:
– Corresponds to EN 13060 or EN 285 and/or
ANSI AAMI ST79
– Suitable programme for the products listed
(e g with hollow bodies, fractionated vacu-
um procedure in three vacuum steps)
– Sufficient product drying
– Validated process in accordance with DIN EN
ISO 17665 (valid IQ/OQ and product-specific
performance appraisal (PQ)
Perform the following steps:
i
Sterilise the parts for sterilisation (at least
20 minutes at 121 °C, at least 4 minutes
at 132 °C or at least 5 minutes at 134 °C)
Do not exceed 138 °C
Marking
i
Mark the packaged, treated medical product
in such a way as to ensure safe application
13.11 Issue clearance for the parts
for sterilisation
The reprocessing of the medical products ends
with the documented clearance for storage and
renewed use
i
Document the clearance of the medical prod-
uct after reprocessing
13.12 Storing parts for sterilisation
i
Comply with the stated storage conditions:
– Store the parts protected against contami-
nation
– Dust-protected, eg in a locked cabinet
– Protected against moisture
– Protected against excessive temperature
fluctuations
– Protected against damage
Packaging for a sterile medical device can
suffer damage as a result of a particular inci-
dent and the passage of time
Potential external contamination of the sterile
barrier system should be taken into account
in terms of aseptic preparation when estab-
lishing the storage conditions
13.9 Packing
CAUTION
Endangering the sterilisation success
The fitted components are not reached
by the steam and as such are not steri-
lised
i
Do not fit the components before
packaging
For packaging of the components, use only
sterile barrier systems made of transparent
paper film that are approved for use in steam
sterilisation according to the manufacturer infor-
mation This includes:
– Temperature resistance up to 138°C
– Standards DIN EN ISO 11607-1/2
– The applicable sections of the standard series
DIN EN 868
The sterile barrier system must be large enough
Once it is loaded, the sterile barrier system must
not be under any strain
13.10 Steam sterilising
WARNING
Health risk due to incorrect sterilisa-
tion
If the sterilisation not performed correct-
ly, it may not be effective The use of in-
struments that have not been properly
sterilised can pose a health risk to the
patient
i
Only steam sterilisation must be used
i
Comply with all of the specified proc-
ess parameters
i
Comply with the manufacturer‘s in-
structions regarding use of the steam
steriliser
i
Do not use any other methods
NOTICE
Damage to equipment due to incor-
rect sterilisation
If the sterilisation process is not per-
formed correctly, this can cause dam-
age to the product
i
Comply with the manufacturer‘s in-
structions regarding use of the steam
steriliser
i
Comply with all of the specified proc-
ess parameters
EN
Содержание Vector Scaler
Страница 1: ...EN 1808V002 9000 615 32 02 0297 Installation and operating instructions Vector Scaler ...
Страница 2: ......
Страница 46: ......
Страница 47: ......