Usage
i
Place the removable parts of the handpiece
(Scaler cover, light conductor), tool kit Scaler
(without instruments) in the ultrasonic bath for
the required reaction time, so that all parts are
covered
i
Clean all accessible external surfaces and in-
ternal surfaces below the surface for 5 min-
utes with a sterile cleaning brush until all visi-
ble contaminants have been removed
i
Place instruments in small parts baskets into
the ultrasonic bath
i
Place the handpiece without covers in a con-
tainer with fluid The drive mechanism of the
handpiece must not lie in the fluid (malfunc-
tion) Therefore observe instructions concern-
ing maximum fluid level for the Scaler hand-
piece
max.
i
Place container with the handpieces in the
ultrasonic bath with a suitable carrier
i
Rinse handpieces through at least 3 times
using a 20-ml disposable pipette
20 ml
i
Screw Scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times using
a 20-ml disposable syringe
20 ml
i
Remove all rinse adapters
i
Comply with the cleaning agents' action times
specified by the manufacturer
Final rinse
After the action time prescribed by the manufac-
turer:
i
Rinse all components under running water for
at least 1 minute (temperature < 20 °C)
Drying
i
If necessary, re-dry at a clean location using a
hygienic, lint-free cloth
i
Blow dry the components with compressed
air in a clean location
13.6 Manual cleaning, intermediate
rinsing, disinfection, final rinse,
drying in ultrasonic bath
A combined cleaning and disinfectant agent is
required for manual cleaning and disinfection It
must have the following properties:
– certified, possibly virucidal efficacy (DVV/RKI,
VAH or European Standards)
– without chlorine, solvents, strong alkaline
solutions (pH >11) or oxidising agents
For further information, see: "General informa-
tion"
Cleaning in an ultrasonic bath
NOTICE
Malfunctions in the handpiece due to
improper handling during cleaning or
disinfecting
i
The handpieces of the Vector must
only be cleaned or disinfected in a
suitable container in an ultrasonic
bath
i
The handpieces must NOT be fully
submerged in fluid