Assembly
i
Open the condensate drain valve and drain
the condensate.
i
Switch on the suction machine by removing
the suction hose from the hose manifold or by
activating the rinsing of the spittoon (spittoon
valve, "wet suction system").
i
Carry out a functional inspection and check
the connections for leaks.
i
For alternating current suction machines,
check the direction of rotation.
The VS suction machines continue after
running for approx. 30 seconds. In case
of simultaneous start signals for the suc-
tion machine and the compressor units,
the compressor units start up with a time
delay to the suction machine.
i
Switch off the suction machine by removing
the suction hose from the hose manifold or by
deactivating the rinsing of the spittoon.
An adequate ventilation by fans in the
unit is only possible when all covers are
attached.
i
Place all covers on the unit.
8.1 Monitoring the device with
Tyscor Pulse
The following units can be monitored with Tys-
cor Pulse:
Power Tower Silence 200/42
surgery supply
0949-200/42
Power Tower Silence 200
surgery supply
A094920035
8 Commissioning
Perform an electrical safety check prior
to commissioning.
i
Main fuses "On"
i
Press the black starting switches on the mo-
tor protection switches.
Only for units with analogue display
i
Adjust the pressure switch to setting "I Auto".
i
Switch on the main power switch.
Compressor will start and switches off at ap-
prox. 0.75 MPa.
EN
Содержание Power Tower Silence 200
Страница 1: ...EN 1806V002 A949100002L02 Installation and operating instructions Power Tower Silence 200 ...
Страница 2: ......
Страница 25: ...A949100002L02 1806V002 23 11 Connection media plan PTS 200 Appendix EN ...
Страница 26: ...24 A949100002L02 1806V002 Appendix PTS 200 PC board mit with CA 4 EN ...
Страница 27: ...A949100002L02 1806V002 25 Appendix EN ...
Страница 33: ...A949100002L02 1806V002 31 Appendix Y1 Relief valve EN ...
Страница 34: ...32 A949100002L02 1806V002 Appendix X3 Top display optional X4 Manifold control contact EN ...
Страница 38: ...36 A949100002L02 1806V002 Appendix M5 Drying unit fan Y1 Relief valve Q4 Pressure switch EN ...
Страница 39: ...A949100002L02 1806V002 37 Appendix X4 24 V AC Ablagesteuerkontakt EN ...
Страница 42: ...40 A949100002L02 1806V002 Appendix Y1 Relief valve EN ...
Страница 43: ...A949100002L02 1806V002 41 Appendix X4 24 V AC manifold control contact EN ...
Страница 46: ...44 A949100002L02 1806V002 Appendix EN ...
Страница 47: ......